Visualisation of lenticulostriate arteries using photon counting detector CT angiography: A retrospective cohort
Ilse Huijberts1, Bart A J M Wagemans2, Florentina M E Pinckaers3
1Department of Radiology and Nuclear Medicine, Maastricht University Medical Centre, Maastricht, the Netherlands; Department of Neurology, Maastricht University Medical Centre, Maastricht, the Netherlands; Mental Health and Neuroscience Research Institute (MHeNs), Maastricht University, Maastricht, the Netherlands; Cardiovascular Research Institute Maastricht (CARIM), Maastricht University, Maastricht, the Netherlands.
Purpose:
The visualisation of lenticulostriate arteries (LSA) using CT-techniques has not been extensively studied. Currently, digital subtraction angiography (DSA) is the preferred imaging modality for LSA visualisation. The new photon-counting detector CT (PCD-CT) may offer a non-invasive alternative. This study evaluates PCD-CT for LSA visualisation compared to DSA and energy-integrating detector CT-angiography (EID-CTA).
Methods:
We assessed imaging data from subarachnoid haemorrhage (SAH) patients, as they are expected to have intact LSA and routinely undergo DSA. From 2021 to 2024, adult SAH-patients who underwent both DSA and either EID-CTA or PCD-CT were enrolled. PCD-CT was acquired in UHR mode. For LSA > 5 mm from the middle cerebral artery (MCA) origin, only the length of the largest branch was measured, whereas for LSA < 5 mm all individual branches were assessed. LSA presence, length and subjective image quality were assessed across PCD-CT, EID-CTA, and DSA by two radiologists. Cohen's kappa and the Intraclass Correlation Coefficient were used to evaluate inter-rater agreement.
Results:
159 patients were included. No differences were found between DSA and PCD-CT in LSA detection and length < 5 mm from the MCA origin. At > 5 mm from the MCA origin, the measured length of LSA was significantly longer on DSA. PCD-CT outperformed EID-CTA for LSA detection and length measurements. Observer 1 rated PCD-CT quality lower than DSA, while observer 2 found no difference. EID-CTA quality was rated significantly lower than both PCD-CT and DSA. Inter-rater agreement ranged from poor to moderate across modalities.
Conclusion:
PCD-CT holds promise as a method for non-invasive LSA visualisation and requires further evaluation in prospective studies and clinical settings.
More Related Videos
13:07Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
06:45Author Spotlight: Integrated Photoacoustic, Ultrasound, and Angiographic Tomography (PAUSAT) for NonInvasive Whole-Brain Imaging of Ischemic Stroke
Published on: June 2, 2023
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...