Related Experiment Video
Updated: Jan 8, 2026

05:31
Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
3.3K
Balloon-Expandable Valve Performance Beyond 10 Years After Transcatheter Aortic Valve Implantation.
Antonin Trimaille1, Marisa Avvedimento1, Pedro Cepas-Guillén1
1Quebec Heart & Lung Institute, Laval University, Quebec City, Quebec, Canada.
The Canadian Journal of Cardiology
|December 20, 2025
Summary
Balloon-expandable valves show good durability after transcatheter aortic valve implantation (TAVI) up to 10 years. While 71% of patients remained free from valve-related issues, some moderate or severe structural valve deterioration was observed.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Limited data exists on the very long-term durability of bioprosthetic valves after transcatheter aortic valve implantation (TAVI).
- Assessing bioprosthetic valve performance beyond 10 years is crucial for understanding long-term outcomes.
Purpose of the Study:
- To evaluate the very long-term clinical efficacy and durability of balloon-expandable valves used in TAVI.
- To identify factors associated with structural valve deterioration (SVD) and valve-related complications.
Main Methods:
- A prospective single-center registry analysis of consecutive patients undergoing TAVI with balloon-expandable valves (2007-2015).
- Focus on patients alive at one-year post-TAVI, with a primary outcome of valve-related long-term clinical efficacy per VARC-3 criteria.
- Death was considered a competing risk in the analysis.
Main Results:
- At 10 years, 71% of patients remained free from the composite endpoint of valve failure, stroke, or bleeding.
- Moderate or severe HVD (structural valve deterioration) occurred in 17% of patients.
- Independent predictors of HVD included younger age, severe patient-prosthesis mismatch, and absence of anticoagulant therapy.
Conclusions:
- Balloon-expandable valves demonstrate reassuring very long-term durability after TAVI, with a majority of patients remaining free from major valve-related events.
- Approximately one-fifth of patients experienced moderate or severe HVD, highlighting the need for ongoing monitoring.
- The potential protective effect of anticoagulation on valve durability warrants further investigation.