Quantification of Scatter Radiation Doses to the Eyes, Thyroid, and Breasts During Plain Abdominal Computed
Kamatchi Krishnamoorthy1, Anil K Sakalecha2, Venkateswarlu Raavi3
1Radiation Oncology, Allied Health Science, Sri Devaraj Urs Academy of Higher Education and Research (Deemed to be University), Kolar, IND.
Purpose:
The use of computed tomography (CT) has increased tremendously, and at the same time, it has also raised concerns about potential risks such as cataracts and cancers among patients and occupational workers. The majority of studies reported the entrance surface dose to the eyes, thyroid, and breast within the CT scan field of view, without shielding. In the present study, we aim to measure the scatter doses (outside the scan area) to the eyes, thyroid, and breast during a plain abdominal CT scan with (0.5 mm lead equivalence) and without shielding.
Materials And Methods:
A cross-sectional analytical study was designed for participants (n = 93; 51 males (55%), 42 females (45%)) and with shielding (n = 93; 52 males (56%), 41 females (44%)) who underwent plain abdominal CT scans. Scatter radiation doses to the eyes, thyroid, and breast were measured using a CaSO₄:Dy Teflon thermoluminescence dosimeter.
Results:
The measured mean ± standard deviation value of without and with shielding are as follows: right eye (0.126 ± 0.170 mSv, 0.030 ± 0.035 mSv), left eye (0.112 ± 0.133 mSv, 0.029 ± 0.035 mSv), right thyroid (0.203 ± 0.296 mSv, 0.102 ± 0.089 mSv), left thyroid (0.210 ± 0.282 mSv, 0.112 ± 0.085 mSv), right breast (0.658 ± 0.863 mSv, 0.287 ± 0.218 mSv), and left breast (0.626 ± 0.842 mSv, 0.273 ± 0.199 mSv). Scatter doses to the eyes were reduced by 74-75% (p < 0.0001), the thyroid by 47-50% (p < 0.0001), and the breast by 56% (p < 0.0001) with shielding compared to scatter doses received by these organs without shielding.
Conclusions:
The present study indicates that organs outside the CT scan area received scatter radiation. Shielding significantly reduced the doses to the eyes, thyroid, and breast. Hence, the findings support the use of shielding whenever feasible during a CT scan.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Biological Effects of Radiation
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...