Growth faltering in a preterm infant associated with inaccurate formula preparation by an automated formula dispenser
Lisa Nobel1, Stefan Johansson2,3, Björn Lundberg2,4
1Sachs' Children and Youth Hospital, Södersjukhuset, Region Stockholm, Stockholm, Sweden lisa.nobel@regionstockholm.se.
Insights
Automated formula dispensers can incorrectly prepare milk, leading to underdilution and faltered growth in infants. Healthcare professionals should improve history-taking and inform caregivers that these devices are not infallible.
Area of Science:
- Pediatrics
- Neonatology
- Infant Nutrition
Background:
- Inadequate caloric intake is a primary cause of faltered growth in infants.
- Proper milk formula preparation, including correct dilution, is crucial for infant development.
- Automated formula dispensers are increasingly used but may pose risks.
Purpose of the Study:
- To report a case of faltered growth in a preterm infant due to underdiluted formula from an automated dispenser.
- To raise awareness among healthcare professionals about the potential risks associated with automated formula dispensers.
- To emphasize the importance of thorough history-taking when evaluating infant faltered growth.
Main Methods:
- Case report of a preterm infant presenting with faltered growth, apnoea, and hypothermia.
- Clinical examination, investigations, and detailed history taking were performed.
- Analysis of milk formula preparation methods, including the use of an automated dispenser.
Main Results:
- The infant's faltered growth was determined to be caused by underdiluted milk formula.
- The underdilution was a result of a malfunction or improper use of an automated formula dispenser.
- The case highlights a potential, previously underestimated, risk in infant feeding technology.
Conclusions:
- Automated formula dispensers, while convenient, can lead to serious feeding errors like underdilution.
- Healthcare providers must remain vigilant in assessing infant feeding practices, even with technological aids.
- Caregivers should be educated that automated dispensers require careful monitoring and are not foolproof.
Abstract:
Inadequate caloric intake is a common cause of faltered growth in infants. When investigating an infant with faltered growth, it is important to ensure recommended feeding volumes. Both overdiluted and underdiluted milk formula are common when mixing milk formula. In recent years, automated formula dispensers have been available to facilitate the preparation of milk formula. Our case illustrates a preterm infant who presented with faltered growth in combination with brief apnoea and hypothermia. After clinical examination, investigation and history taking, it was concluded that the faltered growth was a result of underdiluted milk formula produced by an automated formula dispenser. Our case report aims to raise awareness regarding the potential risk of faltered growth when using an automated milk formula dispenser. Hence, encouraging healthcare professionals to improve the history taking when investigating an infant with faltered growth, as well as informing caregivers that using a milk formula dispenser is not infallible.