Accuracy and reliability of a continuous glucose monitoring system with a focus on hypoglycaemia
Lukas van Baal1, Lorenz Raess1, Annie Mathew1
1Department of Endocrinology, Diabetes and Metabolism, University Hospital Essen, University Duisburg-Essen, Essen, Germany.
Aims:
Glucose monitoring is increasingly based on continuous glucose monitoring (CGM) systems. However, data on the accuracy of CGM within the hypoglycaemic range is sparse. This study investigated CGM accuracy within the hypoglycaemic range in three different clinical settings of hypoglycaemia.
Materials And Methods:
Ninety-two people with various causes of hypoglycaemia were analysed: (i) people during insulin tolerance testing (ITT) (n = 63); (ii) people with insulinoma (n = 16); and (iii) people with diabetes receiving subcutaneous insulin therapy (n = 13). CGM accuracy was evaluated for subgroups and different glucose rates of change (RoC) using mean absolute relative difference (MARD), percentage of glucose values within ±20 mg/dL of point-of-care glucose (%20/20), diabetes technology society error grid and Bland-Altman analysis (BAA).
Results:
Four hundred sixty-four CGM/POC glucose pairs were obtained (39.7% level 1, 60.3% level 2 hypoglycaemia). CGM accuracy decreased from people with diabetes receiving subcutaneous insulin therapy (MARD: 13.9%; %20/20: 81.5%) to those with ITT (MARD: 50.8%; p < 0.01, %20/20: 67.5%; p .02). CGM accuracy significantly decreased with higher RoC. Proportion of CGM values with moderate risk in failing to detect potentially dangerous hypoglycaemia increased from people with diabetes (1.5%) and insulinoma (1.2%) to people with ITT (14.2%, p < 0.01). BAA revealed a significantly increasing bias of -3.3 + 10.7 mg/dL in people with diabetes receiving insulin therapy to -15.2 ± 13.6 mg/dL in people with ITT (p < 0.01).
Conclusion:
CGM accuracy can vary in different clinical hypoglycaemia scenarios. While it remains acceptable in people with diabetes receiving subcutaneous insulin therapy and people with insulinoma, it appears to be inaccurate for glucose monitoring during ITT.
Related Concept Videos
Hypoglycemia and Glucagon
Errors occurring during blood pressure monitoring
Several factors...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...