A proof of concept for improving comparability of dosimetry audits through centralised planning
José Antonio Baeza-Ortega1, Lauren May1, Mohammad Hussein2
1School of Information and Physical Sciences, University of Newcastle, Newcastle, Australia.
Background And Purpose:
The role of dosimetry audits is well established in the development and verification of radiotherapy safety. Differences in planning and beam modelling make inter-centre comparisons challenging, which can be addressed through distribution of centrally created plans. This study developed a centralised planning approach applicable to multiple audit methodologies, using an example of remote patient specific quality assurance assessment, increasing the interpretability of results and facilitating automation and scalability.
Material And Methods:
Starting with an established plan which met all clinical goals, a commercial dose mimicking algorithm was used to replicate this plan to be suitable for multiple treatment machines. Beam and machine limitation data were collected from participating centres to develop universally acceptable beam models. The influence of variation in beam modelling parameters among centres was assessed by creating additional models using the 2.5th, 25th, 75th and 97.5th percentiles of previously reported data. Multi-leaf collimator angle and leaf position, gantry angle and output deviations were then introduced into copies of these plans.
Results:
Introduced delivery errors caused consistent change in dose metrics across machine models (excluding outliers) with a median (range) standard deviation of 1.0 % (from 0.1 % to 1.7 %) demonstrating similar robustness. Beam model variation did not change whether simulated delivery errors were clinically impactful or not for 95 % of tested plans.
Conclusion:
This study lays the foundation for future standardised methodology for dosimetry audits by providing a centralised planning approach that allows a more consistent assessment of centres.
More Related Videos
Related Concept Videos
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Controlled-Current Coulometry: Coulometric Titration
The fundamental requirements for coulometric titrations are (1) 100% efficiency in the reagent-generating electrode reaction and (2) a stoichiometric and preferably rapid reaction between the generated...
Estimation of the Physical Quantities
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Measurement of Air Content in Concrete
The pressure method,...
Design Example: Measuring Distance Between Two Points with Obstructions