Comparison of blood pressure measurements between the Huawei Watch D smartwatch application and the validated Omron
Jorge Velázquez Saornil1, Raúl Frutos Llanes2, Sonia Jiménez Blanco2
1Facultad de Ciencias de la Salud. Departamento de Fisioterapia., Universidad Pontificia de Salamanca, Salamanca, Spain.
Abstract:
Popular wearable devices that record a variety of health metrics, such as real-time blood pressure (BP), could play a role in detecting hypertension in the population. The objective of this study was to compare the validity of the Huawei D device with that of the Omron M3 Intellisense device used by health workers. Sampling was used to obtain patients with normal and abnormal BP. Measurements were taken on the wrist via the Huawei Watch D, while measurements were made using the Omron M3 Intellisense (prevalidated) device used by experienced healthcare personnel at the health centre. A total of 100 normotensive and 100 hypertensive patients were analysed. The concordance statistics were calculated, and linear regression analysis was performed. The intraclass correlation coefficient was calculated to assess the concordance between the instruments used in the measurement and among the observers who made the measurements under a two-factor mixed-effects model, with calculation of the degree of absolute concordance in mean measurements. We found moderate results (systolic blood pressure (SBP) at 30 and 60 min and heart rate at 30 min) and nil for the rest [0.63 (0.36-0.78) < 0.001]; in addition, the comparison of the two groups was verified by a Bland-Altamn graph, in which the disparity of the results was observed. The Huawei D BP measuring device was shown to be relatively effective for measuring BP in hypertensive patients. However, the specificity of BP measurement in normotensive patients was not as reliable as in hypertensive patients.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)