Assessing small-lesion detectability and acquisition time optimisation in silicon-detector-Based PET: a phantom study
Nicholas Leybourne1,2, Vineet Prakash3, Mohammad Hussein4
1Centre for Vision, Speech and Signal Processing, University of Surrey, Stag Hill, Guildford, GU2 7XH, Surrey, Northern Ireland, UK. nl00362@surrey.ac.uk.
Background:
The adoption of silicon photomultiplier (SiPM) detectors over conventional photomultiplier tubes (PMTs) in Positron Emission Tomography (PET) has enhanced overall system performance. In this phantom study, small-lesion detectability was assessed for SiPM-based and PMT-based PET systems for various inhomogeneity sizes, acquisition times and activity contrasts between the inhomogeneity and background.
Methods:
Six spheres of internal diameters ranging between 4.0 mm and 13.0 mm were integrated into a NEMA/IEC PET Body Phantom and filled with fluorodeoxyglucose, with a sphere activity concentration of 29.2 MBq/L and five sphere-to-background activity concentration ratios between 4 and 20. Scans were performed with an SiPM-based system and a PMT-based PET system for each sphere-to-background activity concentration ratio for acquisition times between 1 and 10 min, and image reconstruction was performed with QClear for both systems. Reconstructed images were evaluated for lesion detectability by a lesion detectability index, contrast-to-noise ratio and lesion detectability Likert scales with validation by comparison with the Rose criterion. A model to estimate the acquisition time for each sphere to be detectable was derived and acquisition time was compared.
Results:
The SiPM-based system demonstrated superior lesion detectability, identifying smaller and less active spheres with shorter acquisition times. For a sphere-to-background activity concentration ratio of 10 and a sphere internal diameter of 6.2 mm, the SiPM-based system achieved a contrast-to-noise ratio of 15.8 and a lesion detectability Likert score of 3, compared to 12.0 and 2, respectively, for the PMT-based system. The acquisition time of the SiPM-based system could be reduced by between 1.6% and 89%, depending on sphere size and sphere-to-background activity concentration ratio. The minimum CNR required for a sphere to achieve a detectability Likert score of 0.5 was 6.3, consistent with the Rose criterion.
Conclusion:
SiPM-based PET has enhanced lesion detectability, especially for smaller, less active regions and for shorter acquisition times. A five-point Likert scale is an effective measure of lesion detectability. Guidance is also provided for choosing the acquisition time as a function of lesion size and activity uptake, and for changes in image quality testing protocols.
More Related Videos
14:19A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET