An Interlaboratory Comparison of High-Resolution Ion Mobility Collision Cross-Sections from Human Plasma.

Rachel A Harris1, Emanuel Zlibut1, Allison R Reardon2

  • 1MOBILion Systems, Inc., Chadds Ford, Pennsylvania 19317, United States.

Summary

High-resolution ion mobility (HRIM) measurements show excellent reproducibility for lipid CCS values across laboratories. A new database of 250 lipid collision cross-section (CCS) values supports HRIM adoption in lipidomics.

Related Concept Videos

Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview01:19

Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview

In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
1.7K
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences01:20

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
1.2K
High-Resolution Mass Spectrometry (HRMS)01:15

High-Resolution Mass Spectrometry (HRMS)

The resolution of a mass spectrometer depends on the efficiency of separating ions with different ion masses. The mass of an atom is approximated to the sum of the masses of protons and neutrons inside, considering the masses of protons and neutrons as equal. However, the masses of the proton (1.6726 × 10−24 g) and neutron (1.6749 × 10−24 g) are not truly equal. There is a minor error in the expression of atomic masses relative to the simplest atom of hydrogen. For...
2.3K
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle01:19

Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle

Inductively coupled plasma (ICP) is the most widely used plasma source in atomic emission spectroscopy (AES), also known as Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES). The ICP source, or torch, consists of three concentric quartz tubes with argon gas flowing through them. A spark from a Tesla coil initiates the ionization of argon, generating a high-temperature plasma.
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
1.6K