Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Cuffless blood pressure in 2025: from promise to practice: a narrative review
Farah Wehbe1, Swapnil Hiremath
1Division of Nephrology, Department of Medicine, University of Ottawa and the Ottawa Hospital, Ottawa, Ontario, Canada.
Purpose Of Review:
Cuffless blood pressure (BP) technologies have moved from concept to everyday wearables, spurred by consumer adoption and recent regulatory milestones. Yet clinicians face mixed signals on accuracy, calibration stability, and clinical use. This review synthesizes current evidence and emerging standards to clarify where cuffless data can add value.
Recent Findings:
Most devices estimate BP from surrogate signals and require periodic calibration, making outputs reliable for trends around the last calibration but less so for absolute values. New cuffless-specific validation frameworks mandate dynamic testing (position/hydrostatics, activity/exercise, awake-sleep, pharmacologic response, and prerecalibration drift). Professional bodies currently advise against diagnosis or drug titration using cuffless readings unless a device passes such protocols. Real-world studies show feasibility and patient preference but only moderate agreement with cuffs and limited outcomes data. Special populations (chronic kidney disease/dialysis, pregnancy, pediatrics/frail, diverse skin tones) introduce additional accuracy and equity concerns.
Summary:
Cuffless wearables may be used as adjuncts to surface patterns and support engagement; diagnosis and treatment decisions must be confirmed with validated upper-arm measurements. A pragmatic "good-enough" checklist includes stable calibration over weeks, accuracy across dynamic states and subgroups, and transparent multisite validation. Priorities include harmonized standards, inclusive datasets, and trials linking cuffless monitoring to clinical outcomes.
Related Concept Videos
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.