Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
533
Comparison meta-analysis of two swept-source optical coherence biometers
Daniel Romero1, Rubén Toledano, Alicia Cárceles
1Department of Opthalmology, General University Hospital of Elche, Alacant, Alicante, Spain.
Indian Journal of Ophthalmology
|December 29, 2025
Summary
This study compared two swept-source optical coherence tomography biometers, finding comparable axial length measurements. However, significant differences in white-to-white, lens thickness, and anterior chamber depth may impact intraocular lens power calculations.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Imaging
Background:
- Accurate biometry is crucial for intraocular lens (IOL) power calculation.
- Swept-source optical coherence tomography (SS-OCT) biometers offer advanced measurement capabilities.
- Comparing different SS-OCT devices is essential for clinical practice standardization.
Purpose of the Study:
- To systematically review and meta-analyze the comparability of Anterion and IOLMaster 700 SS-OCT biometers.
- To evaluate differences in key ocular biometric parameters measured by these devices.
- To assess the potential impact of measurement discrepancies on IOL power calculations.
Main Methods:
- Systematic review and meta-analysis of comparative studies published until October 1, 2023.
- Inclusion of 17 studies involving 2943 eyes with cataract or refractive errors.
- Analysis of axial length, keratometry, anterior chamber depth, lens thickness, and white-to-white measurements.
Main Results:
- Axial length measurements were comparable between Anterion and IOLMaster 700 (mean difference -0.01 mm).
- Anterion showed significantly larger anterior chamber depth and lens thickness.
- IOLMaster 700 yielded significantly larger white-to-white measurements; keratometry tended to be higher.
Conclusions:
- While axial length measurements are comparable, significant differences exist in other parameters between Anterion and IOLMaster 700.
- These biometric differences may necessitate adjustments in IOL power calculations when using Anterion data.
- Further research is needed to determine the clinical impact on refractive predictability.