Related Experiment Video
Updated: Jan 7, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
A Clinical Workflow for Evaluating Dose to Organs at Risk After Biology-Guided Radiation Therapy Delivery
Thomas I Banks1, Chenyang Shen1, Andrew R Godley1
1Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX 75390, USA.
Abstract:
Purpose: To develop a workflow for systemically reviewing the doses received by organs at risk (OARs) in a BgRT treatment session, as a means of monitoring delivery constancy and possibly identifying changes of clinical concern in the patient. Methods: We implemented a workflow consisting of a qualitative review of the reconstructed delivered-dose information provided by the RefleXion system, followed by its quantitative evaluation using in-house software. For the latter we developed a framework for calculating custom OAR dose-volume metrics from the delivered-dose distribution and graphing them. We retrospectively applied our workflow to three select BgRT patient cases to appraise its clinical utility. Results: Our workflow efficiently incorporates existing RefleXion TPS features and in-house software into a process for thoroughly evaluating doses to OARs after BgRT delivery. The spreadsheet we created for graphing the trends of normalized OAR dose-volume metrics readily shows if OAR doses have significantly changed or tolerance limits have been violated, thereby potentially revealing if changes of concern occurred in the targeted region. Our workflow also yields a cumulative delivered-dose record at the end of treatment. Conclusions: We established a post-BgRT dose evaluation workflow which supplements the information provided by RefleXion with calculation and graphing of custom OAR dose-volume metrics. This workflow is now routinely used in our clinic following all BgRT treatments. Future improvements could include increased automation, updating the dose calculation to reflect changes in patient anatomy, incorporation of PET metrics, and consideration of target dose data.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...