Challenges in validation studies of blood pressure measurement devices
Hakuo Takahashi1, Kei Asayama2,3
1Department of Cardiology, The Biwako Yoikuin Hospital, Otsu.
Abstract:
Accurate blood pressure (BP) measurement is essential for effective health management. To ensure this, international standards for the validation of BP monitors have been established and revised over time, with ISO 81060-2:2018 being the most recent. While many studies have been published evaluating the accuracy of BP monitors based on this standard, we identified concerns regarding the reliability of some reported results and therefore conducted a comprehensive literature review. Using ISO 81060-2:2018 as the validation framework, we extracted relevant studies from PubMed and collected the SD values for criterion 1 to compare and analyze the reported measurement accuracy. While most SD values ranged from 5 to 8%, we identified six studies reporting values below 4%. On the basis of our experience, such low variability is highly implausible and raises questions about the credibility of these findings. Reports claiming exceptionally high measurement accuracy with unusually low SD values may lead to misconceptions and could potentially be misused for marketing purposes. Therefore, validation studies must be conducted in strict adherence to medical ethics. In addition, there is a need to raise awareness that the current international standards for BP monitor validation require further refinement.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
09:33A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement