Related Experiment Video
Updated: Jun 17, 2026

Video Movement Analysis Using Smartphones ViMAS: A Pilot Study
Published on: March 14, 2017
Quantitative Measurement of Rotational Knee Stability: A Systematic Review of Instrument Reliability and Validity.
Frédérique A C Oosterbaan1, Mariska G H Wesseling1,2, Tom M Piscaer2
1BioMechanical Engineering, Delft University of Technology, Delft, The Netherlands.
Assessing knee stability after anterior cruciate ligament injury requires reliable instruments. This review found no devices with sufficient evidence of both reliability and validity for rotational knee stability, indicating a need for further development.
Area of Science:
- Orthopedics
- Biomechanics
- Sports Medicine
Background:
- Objective evaluation of knee stability in multiple planes offers comprehensive data for anterior cruciate ligament (ACL) injury management.
- Reliability and validity are crucial for assessing the clinical utility of instruments measuring knee stability.
- Existing instruments for rotational knee stability lack clear evidence of adequate reliability and validity.
Purpose of the Study:
- To systematically review instruments measuring rotational knee stability.
- To synthesize existing literature on the reliability and validity of these instruments.
Main Methods:
- A systematic literature search was conducted across four databases.
- Publications reporting reliability or validity of rotational knee stability instruments were included.
- A narrative synthesis approach was used to present findings.
Main Results:
- Twenty-five different instruments were evaluated across 42 studies.
- High heterogeneity was observed in reported parameters and criterion methods.
- Intrarater and interrater reliability were generally adequate (> .75), except under specific conditions (low torque, acceleration/jerk measurement).
- Only 21.1% of validity studies reported very good correlations (r > .8) with criterion measures.
Conclusions:
- No instrument demonstrated sufficient high-quality evidence for both reliability and validity.
- Further research and development are necessary to create and validate devices for objective, multi-planar knee stability assessment.
More Related Videos
04:59Using a Knee Arthrometer to Evaluate Tissue-specific Contributions to Knee Flexion Contracture in the Rat
Published on: November 9, 2018
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Uncertainty in Measurement: Reading Instruments
Quality Assurance