Updated: Jan 7, 2026

Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Activity-dependent transit dose assessment in 192Ir high-dose-rate brachytherapy surface applicator using
Daniel Liu1,2, Emily Draeger3, Christopher J Tien3
1Department of Physics Yale University New Haven USA.
Purpose:
To characterize the newly released GafChromicTM EBT4 film for transit dose measurement in high-dose rate brachytherapy (HDR-BT) surface applicators and to systematically investigate the relationship between transit dose and 192Ir source activity.
Methods:
EBT4 film was used to measure dose at 3 mm depth from a horizontally oriented Leipzig-style surface applicator connected to a 192Ir HDR-BT remote afterloading unit. Transit dose was systematically characterized across eight different source activities ranging from 3.372 to 9.716 Ci (13,724 U to 39,544 U) using measurements with seven nominal dwell times (5-100 seconds, scaled to 10 Ci (40,700 U)). Transit dose was derived from linear extrapolation to zero dwell time and compared with dose calculated (10-Ci source with 100-s nominal dwell time, excluding transit contributions) using the treatment planning system (TPS) Eclipse (Version 16.1).
Results:
EBT4 film demonstrated excellent dose-response linearity with nominal dwell time (adjusted R2 > 0.99) across all source activities, confirming its suitability for transit dose measurement. Transit dose contributions ranged from 0.3% to 4.3% of TPS dose (100-s nominal dwell time at 10 Ci), with maximum contribution from the 9.543 Ci source (0.225±0.040 Gy) and minimum from the 5.502 Ci source (0.017±0.025 Gy). Transit dose showed poor linear correlation with source activity (adjusted R2 = 0.497), indicating that factors beyond source activity influence transit dose magnitude.
Conclusions:
This study validates EBT4 film as a reliable tool for transit dose measurement in surface brachytherapy. The observed non-linear relationship between transit dose and source activity reveals the critical influence of experimental setup variables, particularly source guide tube geometry, which affects the source-to-target distance during transit. These findings demonstrate that transit dose is impacted by multiple interdependent factors including source activity, guide tube configuration, and setup geometry. For clinical applications, this highlights the importance of standardized setup protocols and empirical measurement for accurate transit dose assessment.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Biological Effects of Radiation