Effect of iodinated contrast media enhancement on size-specific dose estimates in emergency thoracic CT
A Khallouqi1, H Sekkat1, A Halimi1
1Sciences and Engineering of Biomedicals, Biophysics and Health Laboratory, Higher Institute of Health Sciences, Hassan First University, Settat, 26000, Morocco.
Introduction:
Accurate estimation of patient radiation dose is crucial in emergency computed tomography (CT), where rapid multiphase imaging is routine. The Size-Specific Dose Estimate (SSDE) incorporates patient size using the water-equivalent diameter (Dw), but standard calculations assume patient size remains constant across scan phases. Contrast administration, however, modifies tissue attenuation and may alter both Dw and SSDE.
Methods:
A retrospective analysis of 100 consecutive emergency thoracic CT examinations was conducted, each including non-contrast (C-) and contrast-enhanced (C+) phases. Automatic tube-current modulation was applied for all acquisitions. From 22,000 slice-level observations, the scanner-reported volume CT dose index (CTDIvol), Dw, size-specific conversion factor (f), and SSDE were calculated using AAPM-recommended equations. Phase-wise comparisons were performed using paired t-tests.
Results:
Mean SSDE increased from 11.64 ± 1.11 mGy (C-) to 11.96 ± 1.17 mGy (C+), a statistically significant 2.7 % rise (p < 0.001). The apparent cause was a 3.4 % decrease in Dw (21.98 mm → 21.24 mm, p < 0.001), reflecting higher post-contrast attenuation interpreted as smaller patient size. This led to a 2.8 % increase in f, further amplified by the scanner's compensatory increase in tube current.
Conclusion:
SSDE is a dynamic rather than static parameter, increasing systematically during contrast-enhanced thoracic CT because of both algorithmic artefacts and real modulation of scanner output. Reliance on a single averaged SSDE across multiphase protocols may underestimate patient dose.
Implications For Practice:
Phase-specific SSDE reporting provides a more accurate and transparent assessment of patient radiation exposure in emergency CT. Integrating automated, phase-resolved SSDE computation into dose-management systems would enhance clinical decision-making, support protocol optimization, and improve adherence to the ALARA principle in daily radiology practice.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...