Performance of continuous glucose monitoring systems (CGMs) during commercial flights in T1DM: A within-subject
Renald Meçani1, Silvia Basta1, Petra M Baumann1
1Division of Endocrinology and Diabetology, Department of Internal Medicine, Medical University of Graz, Graz, Austria.
Aims:
Continuous glucose monitoring (CGM) systems are increasingly used, but their accuracy under flight-conditions remains insufficiently studied. We assessed sensor accuracy of four CGMs during in-flight (mid-haul) and ground conditions.
Materials And Methods:
Participants (n = 20) wore four sensors (Abbott Libre 3, Dexcom G7, Medtronic Simplera, and Sinocare iCan i3) in parallel, inserted 1 day prior to experiment. Capillary reference measurements were performed every 30 min. Accuracy evaluation included: proportion of CGM values within 10% of comparator glucose levels >100 mg/dL or within 10 mg/days of comparator levels ≤100 mg/dL (%10/10), along with %20/20 agreement rates, mean absolute relative difference (MARD), Consensus error grid.
Results:
Twenty participants with type 1 diabetes (age: 37.9 ± 12.8 years, HbA1c: 52.3 ± 4.9 mmol/mol, diabetes duration 19.25 ± 10.6 years) were included in the analysis. Data from 297 sensors with 3473 matched pairs (G7/Libre3/Simplera/i3) were analysed. The overall MARDs were: 9.5% ± 3.3% (G7), 9.6% ± 3.8% (Libre3), 19.9% ± 6.2% (Simplera) and 15.7% ± 6.6% (i3). Overall %20/20 and %10/10 agreement rates were 90.6%/70.2% (G7), 90.8%/66.8% (Libre3), 63.6%/37.8% (Simplera) and 72.5%/42.2% (i3) respectively. The proportion of values in Zones A and B of the Consensus error grid was 98.9% (G7), 99.0% (Libre 3), 94.5% (Simplera) and 97.5% (i3). Prandial state and flight phase had limited influence on the relative performance ranking of the devices.
Conclusions:
G7 and Libre3 showed consistently higher accuracy and clinical reliability across various phases of flight and ground conditions. Simplera and i3 exhibited greater variability in accuracy. These findings suggest that while modern CGMs could support diabetes management in aviation, individual system performance may differ under specific operational conditions.
Related Concept Videos
Diabetes Mellitus: Type 2 and Gestational
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Hypoglycemia and Glucagon
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...