Facile Preparation of Ultrafine Aluminum Hydroxide Particles with or without Mesoporous MCM-41 in Ambient Environments
Published on: May 11, 2017
Beam quality assessment: Evaluating aluminum purity's effects.
Sara Mohammadi1, Joshua Deslongchamps1, Jiali Wang1
1Department of Medical Imaging Technology and Informatics, Southern California Permanente Medical Group, Los Angeles, California, USA.
Aluminum purity significantly impacts Half Value Layer (HVL) measurements, affecting radiographic beam quality assessments. Higher aluminum purity leads to increased HVL values, potentially causing non-compliance with regulatory standards.
Area of Science:
- Medical Physics
- Radiological Sciences
- Radiation Protection
Background:
- Beam quality assessment is crucial for safe and effective ionizing radiation delivery in medical imaging.
- Variations in national and international standards for beam quality assessment can lead to measurement ambiguities and regulatory compliance issues.
Purpose of the Study:
- To evaluate the impact of aluminum (Al) purity on Half Value Layer (HVL) measurements and radiographic beam quality.
- To assess these effects across both mobile and fixed radiographic units.
Main Methods:
- Utilized aluminum filters of 99.0% and 99.5% purity for beam filtration measurements.
- Followed International Electrotechnical Commission (IEC) standards for HVL measurements using radiation detectors from three manufacturers.
- Determined HVL through exponential fitting of exposure versus aluminum thickness and performed single-shot HVL measurements.
Main Results:
- Increasing aluminum purity from 99.0% to 99.5% resulted in a 1.35% to 4.82% rise in HVL measurements (median increase of 3.26%).
- Deviation between single-shot and interpolated HVL measurements varied, with higher deviations observed for 99.0% purity aluminum.
- Results demonstrate a clear sensitivity of HVL measurements to aluminum purity.
Conclusions:
- A discrepancy exists between FDA (≥99.0% purity) and IEC (≥99.9% purity) standards for reference aluminum purity.
- Variations in aluminum purity can lead to radiographic units falsely failing minimum HVL regulatory requirements.
- Medical physicists must be aware of aluminum purity's influence on HVL testing for regulatory compliance.