Mechanically-foldable axial flow blood pump: response-surface-based structural optimization and hemolytic performance
Tairan Ji1, Teng Jing1, Jianan Cheng1
1National Research Center of Pumps, Jiangsu University, Zhenjiang, Jiangsu, China.
This study introduces a foldable blood pump designed to reduce hemolysis by allowing lower rotational speeds. The optimized design improves head pressure by 6.9% and reduces hemolysis by 17.9%.
Area of Science:
- Biomedical Engineering
- Cardiovascular Devices
- Fluid Dynamics
Background:
- Traditional percutaneous ventricular assist devices (PVADs) require high rotational speeds, leading to significant hemolysis due to elevated shear stress.
- Design limitations of existing PVADs necessitate improvements for enhanced clinical applicability and reduced blood trauma.
Purpose of the Study:
- To design and optimize a foldable axial-flow blood pump with improved hydraulic efficiency and reduced hemolysis.
- To systematically analyze the influence of structural parameters on pump performance and blood trauma using numerical simulations.
Main Methods:
- Utilized numerical simulations and response surface methodology (RSM) with Box-Behnken Design (BBD) for parametric optimization.
- Analyzed the impact of key parameters including impeller and diffuser angles, and impeller-diffuser gap on pump head and hemolysis index (HI).
Main Results:
- Identified five critical structural parameters influencing pump head and HI.
- Achieved a 6.9% increase in head pressure and a 17.9% reduction in HI compared to the initial design through optimization.
- Demonstrated that increasing tip clearance reduces hemolysis with a moderate decrease in head.
Conclusions:
- Developed and optimized a foldable axial-flow blood pump for transaortic implantation with enhanced performance.
- The optimized pump configuration achieved a head of 2.346 m and an HI of .
- Further adjustments in tip clearance can offer additional hemolysis reduction without significantly compromising pressure head, indicating potential for further clinical refinement.