Related Experiment Video
Updated: Jan 13, 2026

Engineering Platform and Experimental Protocol for Design and Evaluation of a Neurally-controlled Powered Transfemoral Prosthesis
Published on: July 22, 2014
Investigation and comparison of microprocessor -controlled knee prostheses used in adult people with amputations: A
Fatma Özcan1, Ayşe Naz Kalem Özgen1, Mustafa Hilmi Pekalp2
1Department of Physical Medicine and Rehabilitation, University of Health Sciences, Gaziler Physical and Rehabilitation Medicine Training and Research Hospital, Ankara, Turkey.
Abstract:
The aim of the study was to analyze and compare data obtained from different types of microprocessor-controlled knee prostheses that failed in amputated people with transfemoral and knee disarticulation. The medical records of adult amputees whose microprocessor-controlled knee joint prosthesis failed were reviewed. The Genium X3 had a statistically significant higher daily prosthesis use time than the Plié 2.0 and Plié 3.0 groups (p = 0.033), Orion (p = 0.009) and C-Leg 3 (p = 0.017). Patients who used the Orion, C-Leg 4, Plié 2.0 and Plié 3.0 prostheses had a statistically significantly longer duration of use of the current prosthesis than those who used the Genium, and patients who used the C-Leg 3 had a statistically significantly longer duration of use than those who used the Genium X3 (p = 0.023, p = 0.042, p = 0.010, p = 0.018, respectively). There was a statistically significant correlation between current age and time since amputation and duration of current prosthesis use (p = 0.001, p < 0.01, respectively) (r = 0.315, r = 0.364, respectively). There was a significant correlation between time since amputation and daily prosthesis use time (p = 0.015, r = -0.253). In adult transfemoral and disarticulated knee people with amputations, daily prosthesis use time decreases and current prosthesis use time increases with increasing amputation duration.