Effect of tightening protocols on abutment screw loosening and morphological changes in two-piece zirconia implants:
Toma Hara1, Takayuki Sugiki2, Munetsugu Seto2
1Functional Occlusal Treatment, Department of Crown and Bridge Prosthodontics, The Nippon Dental University Graduate School of Life Dentistry at Niigata, Niigata, Japan.
Purpose:
This study evaluates the effects of different abutment screw tightening protocols and implant materials on screw loosening and morphological changes after cyclic torsional loading of two-piece implants.
Methods:
Titanium (T) and zirconia (Z) two-piece implants were used. Abutment screws were tightened to 35 Ncm. Three protocols were tested: (A) retightening after 5 min, (B) immediate retightening, and (C) no retightening (n = 6 each). All samples underwent 100,000 cycles of cyclic torsional loading. Removal torque values before and after loading (RT-pre and RT-post) and torque loss percentage (TL) were calculated. Surface morphology was assessed using scanning electron microscopy (SEM), and dimensional changes were quantified using root mean square (RMS) values from three-dimensional data.
Results:
In the titanium group, protocol A (TA) had lower RT-post values than protocols B (TB) and C (TC) (P < 0.001), and TB was lower than TC (P < 0.05). In the zirconia group, protocol A (ZA) was significantly lower than protocols B (ZB) and C (ZC) (P < 0.05, P < 0.001, respectively). RT-post and RT-pre were significantly lower in the zirconia group than in the titanium group under all conditions (P < 0.001). No differences were observed in TL. SEM revealed wear on the screw heads and threads in all groups. RMS values were significantly lower in the zirconia group (P < 0.05).
Conclusions:
Abutment screw retightening does not consistently enhance torque maintenance under torsional loading. The implant material affects torque retention and deformation, where titanium implants demonstrating greater stability than zirconia implants.
More Related Videos
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
Related Concept Videos
Normal Strain under Axial Loading
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...