Objective:
Grasping force control is crucial for safe laparoscopic surgery. However, force feedback is limited as haptic information on grasping strength and tissue stiffness is mostly lost due to internal instrument backlash and friction. This increases tissue trauma risk as excessive grasping forces can lead to (postoperative) complications. This study aims to develop a grasping force feedback providing add-on for a laparoscopic grasper and to validate its impact on skills acquisition in basic laparoscopic skills training.
Method:
The ShaftFlex, a shaft-based grasping force measurement system providing feedback was designed as an add-on for standard reusable instruments. It consists of a compliant element deflecting proportionally to the applied grasping force, and a Hall sensor measuring that deflection. Influence on skills acquisition was evaluated in a comparative study where novices were divided into a Feedback and No feedback group, performing five training trials of a silicon torus transfer boxtrainer task. Afterwards, both groups performed a post-training task without feedback. Grasping force, time to completion and number of errors were measured.
Results:
There was a significant difference in mean grasping force between groups for all training trials and the post-training trial. In the Feedback group, there was no significant increase in grasping force when feedback was removed.
Conclusion:
The ShaftFlex working principle provided a feasible, sustainable method to measure grasping forces exerted by a laparoscopic grasper, enabling immediate haptic feedback. It potentially enhances objective skill assessment, providing feedback on training performance. In a clinical context, the ShaftFlex might be useful in surgery where delicate tissue is grasped.
Related Concept Videos
Assessment of the Abdomen III: Palpation
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...