Establishment of allowable total error for semen analysis based on the state of the art in China
Yan Zheng1,2, Lin Yu1,2, Yingbi Wu1,2
1Department of Andrology/Human Sperm Bank, West China Second University Hospital, Sichuan University, Chengdu, China.
Objectives:
To establish evidence-based analytical performance specifications (APS), specifically allowable total error (TEa), for five fundamental semen analysis parameters (sperm concentration, progressive motility, total motility, morphology, and vitality) in China utilizing the state-of-the-art model, and to compare these specifications with existing international benchmarks.
Methods:
We examined national external quality assessment (EQA) data from 2019 to 2025, encompassing 315 laboratories across 30 provinces in China. TEa was calculated using the state-of-the-art model as half the 10th-90th percentile range of result deviations from assigned values. Given that the participant count was below 100 from 2019 to 2022, this study eventually utilized data from 2023 to 2025 for the calculation of TEa. The pass rate of participating laboratories was evaluated against the established TEa criteria.
Results:
The established TEa values as follows: sperm concentration, 34.3 %; progressive motility, 26.9 % (for assigned value ≤40 %) and 13.5 % (>40 %); total motility, 26.3 % (≤50 %) and 10.5 % (>50 %); morphology, 67.7 %; vitality, 24.6 % (≤60 %) and 7.9 % (>60 %). The pass rates for all parameters in recent EQA cycles were approximately 80 %, supporting the practical achievability of the proposed specifications. Comparisons showed that the derived TEa values are generally consistent with or more refined than previous state-of-the-art and biological variation-based standards.
Conclusions:
This study provides the first large-scale, evidence-based TEa for semen analysis in China. The stratified TEa for motility and vitality based on clinical thresholds, improves diagnostic accuracy. The proposed specifications are practical, achievable, and clinically relevant, providing a consistent foundation for quality assurance in reproductive medicine laboratories.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Data Validation
Key parameters for method validation include:
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...