Related Experiment Video
Updated: Jun 16, 2026

Assessment of Bone Fracture Healing Using Micro-Computed Tomography
Published on: December 9, 2022
Photon-Counting CT for Diagnosing Vertical Root Fractures in Teeth With Metal Posts: An Ex Vivo Comparative Analysis
Renata M S Leal1, Fernanda B Fagundes2, Maria F S A Bortoletto2
1Department of Restorative Dentistry, Federal University of Paraná, Paraná, Brazil.
Objective:
Photon-counting computed tomography (PCCT) represents a major innovation in X-ray detection technology, offering improved signal efficiency and reduced electronic noise compared with cone-beam computed tomography (CBCT), which can enhance image quality. This study aimed to evaluate the diagnostic performance of PCCT in detecting vertical root fractures (VRF), in comparison with four CBCT devices.
Methodology:
Eighteen single-rooted teeth were endodontically treated, and VRF was induced in eight of them. Each tooth was individually placed into the mandibular first premolar empty socket of an anthropomorphic phantom and scanned under three conditions: without a metal post, with a nickel-chromium metal post (Ni-Cr), and with a cobalt-chromium metal post (Co-Cr) in five CT devices: the NAEOTOM Alpha PCCT (Siemens Healthineers) device and four CBCT devices (3D Accuitomo 170-Morita, Veraview X800-Morita, NewTom VGi evo-NewTom, and Carestream 9600-Carestream). The highest-resolution protocol available on each device was used, resulting in a total of 270 scans. Five experienced dentomaxillofacial radiologists independently and blindly evaluated the scans using a five-point confidence scale. Diagnostic accuracy was assessed by calculating the area under the ROC curve (AUC), sensitivity, and specificity, with results compared by two-way ANOVA with post hoc Tukey's test (α = 0.05).
Results:
NewTom VGi and PCCT devices showed significantly higher AUC values than the Veraview X800, regardless of the metal post material (p < 0.05). CS9600 and PCCT devices exhibited significantly higher sensitivity values in diagnosing with Ni-Cr posts than the Accuitomo 3D and Veraview X800 devices (p < 0.05). With the Co-Cr metal post, the NewTom VGi, CS9600, and PCCT devices showed significantly higher sensitivity values compared to the Veraview X800 device (p < 0.05). There were no statistically significant differences in specificity, regardless of the CT device or metal post material (p > 0.05).
Conclusions:
The NEAOTOM Alpha PCCT showed high diagnostic accuracy for VRF detection in an ex vivo model, comparable to high-resolution CBCT devices, highlighting its diagnostic performance under controlled ex vivo conditions.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...