Exploring the application of various condenser microphones for wrist pulse measurement using machine learning models
Chetna Sharma1,2, Neha3, Gurinderjit Singh1
1CSIR-Central Scientific Instruments Organisation, Chandigarh, 160030, India.
Unidirectional condenser microphones offer superior accuracy for wrist pulse measurement compared to omnidirectional types. Machine learning models, particularly Nu-Support Vector Regression, enhance pulse rate prediction accuracy.
Area of Science:
- Biomedical Engineering
- Cardiovascular Health Monitoring
- Sensor Technology
Background:
- Wrist pulse measurement is vital for cardiovascular health assessment.
- Existing sensors (optical, pressure, image, ultrasonic) face limitations like environmental interference and system complexity.
- Condenser microphones are emerging as a viable alternative for pulse detection.
Purpose of the Study:
- To evaluate the effectiveness of different condenser microphone types for wrist pulse measurement.
- To compare the accuracy of omnidirectional versus unidirectional condenser microphones.
- To assess the performance of machine learning models in predicting pulse rate using microphone data.
Main Methods:
- Three condenser microphones (one omnidirectional, two unidirectional) were tested.
- A mechanical pulse acquisition system was developed.
- Data from 27 healthy subjects were analyzed using four regression models (Partial Least Square Regression, Ridge Regression, Principal Component Regression, Nu-Support Vector Regression).
Main Results:
- Unidirectional condenser microphones demonstrated higher accuracy than the omnidirectional type.
- A unidirectional microphone with a sensitivity of -50 to -44 dB outperformed one with -40 to -34 dB.
- Nu-Support Vector Regression achieved the lowest prediction errors, showing superior performance.
Conclusions:
- Unidirectional condenser microphones are more suitable for wrist pulse measurement.
- Microphone sensitivity plays a crucial role in accuracy.
- Machine learning, especially Nu-Support Vector Regression, significantly improves pulse rate prediction accuracy, guiding future sensor development.
Related Concept Videos
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Measurement of Blood Pressure