Related Experiment Video
Updated: May 3, 2026

Atomic Force Microscopy of Red-Light Photoreceptors Using PeakForce Quantitative Nanomechanical Property Mapping
Published on: October 24, 2014
Research on the method for measuring the focal spot size of micro-focus X-ray sources using the JIMA resolution test
Li Fengxiao1,2, Wang Yixin1,2, Xu Haodong1,2
1Key Laboratory of Optoelectronic Technology and Systems of the Ministry of Education, ICT Research Center, Chongqing University, Chongqing 400044, China.
Abstract:
BackgroundMeasuring an X-ray source's focal spot size is vital for Micro-CT resolution. Standard methods are often too complex or inaccurate. The popular JIMA resolution test card is simple to use but lacks a clear, quantitative formula to determine the actual focal spot size.ObjectiveThis study aims to create a reliable quantitative link between JIMA resolution and focal spot size using simulations and experiments.MethodsWe used Monte Carlo simulations and practical experiments to establish the relationship between JIMA resolution and focal spot size.ResultsWe found that the focal spot size is twice the line pair width on the JIMA card when the image contrast (MTF) is at 10%. This method is highly accurate, with a maximum measurement error of less than 8.7% compared to a high-precision technique.ConclusionsOur findings provide a simple, fast, and validated method for measuring focal spot size using the JIMA test card. This makes it a practical and reliable alternative to more complex procedures.
More Related Videos
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Confocal Fluorescence Microscopy
Super-resolution Fluorescence Microscopy
X-ray Imaging