Bridging the analog divide: a comparison of printed X-ray films and digital images when using computer-aided
Andrew J Codlin1,2, Thang P Dao3, Binh H Nguyen3
1Friends for International TB Relief, Ha Noi, Viet Nam. andrew.james.codlin@ki.se.
Background:
Computer-aided detection (CAD) software provides scalable, standardized chest X-ray (CXR) interpretation, helping address the global shortage of radiologists and inter-reader variability. Printed X-ray films remain common in many low-resource settings, yet most CAD software can only process Digital Imaging and Communications in Medicine (DICOM) files. Genki software (DeepTek, India) is one of the few World Health Organization (WHO)-recommended CAD software capable of interpreting both DICOM files and photographs of printed X-ray films (Joint Photographic Experts Group [JPEG] files), but its performance using JPEG files has not been independently evaluated.
Methods:
We evaluated Genki software using a test library of 1466 CXR images from adults screened for tuberculosis (TB) in Ho Chi Minh City, Viet Nam. Each participant's TB status was determined using a composite reference standard, based on radiological findings and Xpert MTB/RIF Ultra testing. Each CXR image was blindly re-read by 10 human readers and processed by Genki software using both DICOM and JPEG files. Genki software performance was evaluated using median abnormality scores, area under the receiver operating characteristic curves (AUC), and sensitivity/specificity comparisons at different abnormality score thresholds.
Results:
Genki software abnormality scores were significantly higher when using JPEG files, but this did not translate into significant differences in AUCs between the file types (DICOM AUC = 0.94 vs JPEG AUC = 0.92, p = 0.190). When abnormality score thresholds were calibrated to match average human reader sensitivity (79.0%), Genki achieved significantly higher specificity with both DICOM (95.2% vs 84.8%, p < 0.001) and JPEG (92.1% vs 84.8%, p < 0.001) files. When the software's abnormality score thresholds were calibrated to achieve 90% sensitivity, Genki maintained high specificity with both DICOM (89.3%) and JPEG (81.1%) file types, meeting the minimum Target Product Profile (TPP) criteria for a high-sensitivity, high-specificity screening test.
Conclusions:
Genki software performs comparably when interpreting DICOM and JPEG files, outperforming human readers and meeting TPP criteria with both file types. This capability enhances its usability in resource-limited settings where digital infrastructure is lacking, supporting its broader deployment for TB screening. Further research is needed to assess real-world implementation feasibility and performance in diverse populations and clinical environments.
Related Concept Videos
Pulmonary Tuberculosis IV
Several diagnostic approaches are used to detect TB. The conventional method is the Tuberculin Skin Test (TST), also known as the Mantoux test. However, this method has...
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...