Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Non-invasive blood pressure monitoring using wearables for cardiovascular risk assessment: a systematic review
Michelle Zwahlen1, Elena Pavicic2,3, Kerstin Khattab4
1Faculty of Medicine, University of Bern, Bern, Switzerland.
Insights
Wearable devices show potential for continuous blood pressure monitoring in women, aiding cardiovascular risk assessment. However, accuracy and standardization challenges remain for widespread clinical use.
Area of Science:
- Cardiovascular medicine
- Biomedical engineering
- Preventive healthcare
Background:
- Cardiovascular diseases (CVDs) are a leading cause of death in women globally.
- Hypertension is a significant risk factor for CVDs.
- Wearable devices offer non-invasive, continuous blood pressure (BP) monitoring as an alternative to traditional cuff-based methods.
Conclusions:
- Wearable BP monitoring technology is advancing, offering potential in women's preventive cardiovascular medicine.
- Clinical implementation is hindered by inconsistent accuracy, calibration needs, and lack of standardization.
- Further validation and longitudinal studies are essential to establish their role in cardiovascular risk prediction.
Purpose:
Cardiovascular diseases are the leading causes of mortality in women worldwide, with hypertension being a major risk factor. While traditional blood pressure monitoring techniques rely on cuff-based measurements, wearable devices offer a promising alternative for continuous and non-invasive blood pressure tracking. This systematic review investigates the extent to which wearable blood pressure measurements can serve as surrogates for traditional sensors to be implemented in risk assessment tools in predicting cardiovascular risk in women.
Methods:
A systematic search was conducted in databases including MEDLINE, Embase, Cochrane Library, Web of Science, Scopus and ClinicalTrials.gov. Studies published between 2010 and 2024 were included. Exclusion criteria were case reports or animal studies. Study selection was performed based on PRISMA guidelines. Data extraction focused on wearable devices, measurement approach, validation against gold-standard BP methods, and their predictive utility for cardiovascular outcomes.
Results:
The systematic literature search revealed 14'863 results after removal of duplicates, of which 245 were selected for extraction. Most included studies used photoplethysmography, with pulse transit time and pulse wave velocity as core parameters, some using machine learning. Accuracy was moderate to high for diastolic blood pressure, but systolic blood pressure showed greater variability. Cardiovascular risk stratification showed promising results, though external validation was rare.
Conclusion:
Wearable blood pressure monitoring technologies are maturing, with growing potential in preventive cardiovascular medicine in women. However, clinical implementation is limited by varying accuracy, need for calibration, and the lack of standardization. Further validation and longitudinal studies are needed to establish their role in cardiovascular risk prediction.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Errors occurring during blood pressure monitoring
Several factors...