Protocol development for high-resolution transcutaneous CO₂ monitoring in ultra-early detection of free flap
Ryutaro Shibata1, Toshiya Kudo2, Shinsuke Takeda3
1Trauma Reconstruction Center, Shinyurigaoka General Hospital, Kawasaki, Japan; Department of Orthopaedic Surgery, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan; Shizuoka Graduate School of Public Health, Shizuoka Graduate University of Public Health, Shizuoka, Japan.
Background:
Vascular compromise is a serious complication in free-flap surgery for traumatic reconstruction or fracture-related infections, often leading to partial or total flap loss if not detected promptly. We evaluated transcutaneous partial pressure of carbon dioxide (TcPCO₂) real-time monitoring as an objective, non-invasive method for ultra-early detection of vascular compromise in free flap reconstruction.
Methods:
This sequential cohort study consisted of a retrospective development phase and a prospective validation phase. An abnormality was defined as a rise of >10 mmHg from the baseline TcPCO₂ value, with re-exploration performed if the elevation persisted after recalibration. High-resolution (1-s interval) TcPCO₂ data were analyzed to assess diagnostic accuracy and concordance with arterial partial pressure of carbon dioxide (PaCO₂).
Results:
In pilot studies, TcPCO₂ increased within 20-100 s of induced ischemia and correlated strongly with PaCO₂ (r = 0.708, p < 0.001). Among 81 clinical free flap cases (50 retrospective, 31 prospective), TcPCO₂ monitoring detected all seven episodes of vascular compromise with 100% sensitivity and specificity, and no false positives. All the compromised flaps were successfully salvaged. Compared with conventional clinical assessment, TcPCO₂ monitoring provided earlier recognition of perfusion disturbances.
Conclusion:
TcPCO₂ monitoring is a non-invasive, objective, and reproducible tool for ultra-early detection of vascular compromise in free flap surgery. Its implementation enables timely re-exploration, reduces reliance on subjective bedside assessments, and may significantly improve flap salvage outcomes.
Related Concept Videos
08:18Engineered Vascularized Muscle Flap
06:27Transcutaneous Microcirculatory Imaging in Preterm Neonates
07:18Transcutaneous Assessment of Renal Function in Conscious Rodents
08:45Cerenkov Luminescence Imaging (CLI) for Cancer Therapy Monitoring
07:35Use of Ultra-high Field MRI in Small Rodent Models of Polycystic Kidney Disease for In Vivo Phenotyping and Drug Monitoring
03:19Transcutaneous Neuromuscular Electrical Stimulation for Treating Varicocele-Induced Scrotal Pain