Established acetabular radiological reference values can be reliably transferred to reconstructed parallel-beam 2D
Arnaud Klopfenstein1, Thomas Marth2,3, Stefan Sommer3,4
1Orthopaedic Department, Balgrist University Hospital, University of Zurich, Forchstrasse 340, 8008, Zurich, Switzerland.
Background:
The anteroposterior radiograph of the pelvis is essential for diagnosing hip pathologies. Radiograph-like projections reconstructed as cone-beam images from CT data correlate strongly with conventional radiographs. However, CT inherently uses parallel x-rays rather than a cone-beam geometry. This study aims to determine whether parallel-beam radiograph-like projections from CT provide comparable reference values to cone-beam reconstructions.
Methods:
63 patients (126 hips) undergoing CT for symptomatic hip pathologies without prior hip surgery were included. From the same CT data, cone-beam and parallel-beam radiograph-like images were reconstructed using a standardized algorithm. Reference values, including lateral center-edge angle (LCEA), medial center-edge angle (MCEA), acetabular index (AI), acetabular arc (AA), extrusion index (EI), crossover sign, and posterior wall sign, were measured on both projection types, eliminating technical bias. Two observers performed all measurements twice to assess inter- and intra-observer reliability, and intraclass correlation coefficients (ICCs) were calculated.
Results:
126 hips were analyzed: 52 with LCEA <22° ("acetabular undercoverage"), 49 with LCEA 22°-33° ("normal coverage"), and 25 with LCEA >33° ("acetabular overcoverage"). ICCs between observers and between projection types demonstrated good to excellent reliability for all reference values (0.89-0.99).
Conclusion:
Parallel-beam radiograph-like projections demonstrate good to excellent reliability (ICCs: 0.89-0.99) for key reference values of hip pathologies compared to cone-beam radiograph-like projections from the same CT data. These findings suggest parallel beam projections can be reliably used with established reference values for conventional radiographs.
More Related Videos
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
11:45Functional Interrogation of Adult Hypothalamic Neurogenesis with Focal Radiological Inhibition
Published on: November 14, 2013
Related Concept Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Radiological Investigation I: X-ray and CT
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
11:45Functional Interrogation of Adult Hypothalamic Neurogenesis with Focal Radiological Inhibition
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice
Micro-CT Imaging of a Mouse Spinal Cord
It's a little-known fact that the discovery and (inadvertent) use of X-rays garnered the first ever Nobel Prize in Physics. The famous X-ray image of Dr. Röntgen's wife's hand from 1895 that sent shock waves through the scientific community looks like most modern day 2D medical X-ray images. Though it is not the newest technology,...