Attenuation-based ultra-low-dose lung computed tomography at 0.1 mSv to 0.3 mSv effective dose in children
Max-Johann Sturm1,2, Christian J Kellenberger1,3, Franco Rupcich4
1Department of Diagnostic Imaging, University Children's Hospital Zurich, Lenggstrasse 30, Zurich, 8008, Switzerland.
Background:
Radiation dose reduction is essential in paediatric lung computed tomography (CT). Advances in energy-integrating detector CT and deep-learning reconstruction may enable ultra-low-dose imaging comparable to photon-counting CT.
Objective:
To evaluate the radiation dose and performance of an ultra-low-dose lung CT protocol using a wide-detector energy-integrating CT system in paediatric patients, focusing on effective radiation dose and diagnostic image quality.
Materials And Methods:
A total of 277 low-dose lung CT scans from 106 paediatric patients (age range, 113 days to 17.85 years) were retrospectively analysed. All scans were acquired in axial mode using a 256-slice-multidetector CT scanner with deep learning image reconstruction and attenuation-based Auto Prescription. Radiation dose parameters, including volume CT dose index, dose-length product, size-specific dose estimate, and effective dose, were calculated. Signal-to-noise ratio and contrast-to-noise ratio were assessed in standardised anatomical regions. Patients were stratified by age, and statistical analysis was conducted to evaluate dose trends and image quality metrics.
Results:
There were significant differences between all age groups for all dose parameters (Kruskal-Wallis test, P<0.05). The median effective dose increased with age, ranging from 0.12 mSv (interquartile range (IQR) 0.09-0.14 mSv) in the 0-5-year group to 0.23 mSv (IQR 0.21-0.25 mSv) in adolescents aged 15 years to <18 years. Contrast-to-noise ratio and signal-to-noise ratio exhibited age-dependent variation with a small increase in older age groups. One-sided non-inferiority testing demonstrated that the signal-to-noise ratio and contrast-to-noise ratio in the youngest age group (0-5 years) were not significantly inferior to those in the ≥15-year group (P<0.05). All examinations were deemed diagnostically sufficient by board-certified paediatric radiologists. Non-disruptive artefacts such as cardiac motion and step artefacts occurred frequently but did not impair interpretation.
Conclusions:
Ultra-low-dose lung CT using wide-detector energy-integrating CT with deep-learning image reconstruction allows for routine diagnostic imaging in children at radiation doses ranging from 0.12 mSv to 0.23 mSv, comparable to those reported for newer photon-counting CT systems. This approach provides a robust, clinically viable strategy for minimizing radiation exposure while maintaining diagnostic image quality.
Related Concept Videos
Drug Dosing: Infants and Children
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant