Pedicle Screw Placement Using an Augmented Reality Head-Mounted Display in a Porcine Model
Published on: May 24, 2024
Early Clinical Safety Profile and Accuracy of Augmented Reality-Assisted Pedicle Screw Placement: A Multi-Center
Luis M Salazar1, Rajko S Vucicevic1, Justin B Castonguay1
1Department of Orthopaedic Surgery, Rush University Medical Center, Chicago, IL.
Study Design:
Retrospective case series.
Objective:
To report on the early clinical safety profile and accuracy of Augmented reality (AR)-assisted pedicle screw placement in patients undergoing minimally invasive spinal surgery.
Summary Of Background Data:
While navigation-guided pedicle screw placement has improved accuracy rates to 89-100%, current systems face challenges, including line-of-sight interruptions and attention shift from patient to remote screens. AR technology projects navigation data directly onto the surgical field, potentially overcoming these limitations while maintaining high accuracy.
Methods:
A multi-center retrospective analysis was conducted across two institutions, reviewing 31 procedures involving 200 pedicle screws placed using the Xvision AR navigation system (Augmedics Inc.). All procedures were performed by fellowship-trained spine surgeons. Accuracy was assessed using pre- and post-instrumentation CT O-arm imaging according to the Gertzbein and Robbins classification system. Novel angular deviation analysis was performed to evaluate trajectory accuracy. Clinical outcomes, including complications and revision rates, were recorded.
Results:
Among 200 pedicle screws analyzed, 84.5% were classified as Gertzbein-Robbins grade A (complete containment), 13.5% as grade B (trivial breach), and 2% as grade C, yielding an overall accuracy rate of 98%. The pedicle breach rate was 15.5%, with no instances of superior endplate or facet joint violations. Mean angular deviation from the ideal trajectory was 0.78±4.56 degrees. No screw revisions, hardware-related complications, or symptomatic nerve encroachment occurred. Two patients experienced non-hardware-related systemic infections.
Conclusions:
AR-assisted navigation demonstrates high accuracy for pedicle screw placement with minimal angular deviation and low complication rates. This technology offers promising potential for improving surgical precision while maintaining workflow efficiency.
Level Of Evidence:
Level IV.
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Survey Safety