Enhanced Fatigue Performance of VAR/EBR Nitinol for Transcatheter Heart Valve Applications: Material Characterization
Maximilien E Launey1, Trey Bobo2, Behnood Miri2
1G.RAU Inc., Scotts Valley, California, USA.
None:
Transcatheter cardiovascular devices require high-purity Nitinol materials with exceptional fatigue resistance to meet stringent Class III regulatory durability requirements. Ultra-clean VAR/EBR (Vacuum Arc Remelt/Electron Beam Remelt) Nitinol represents a metallurgical advancement that achieves unprecedented control over inclusion size and distribution. This study characterizes the fatigue behavior of VAR/EBR Nitinol with nominal inclusion sizes below 10 μm under conditions representative of transcatheter mitral valve replacement (TMVR) applications, using the HighLife Mitral Valve Replacement system as a clinical case study. Diamond-shaped fatigue specimens were manufactured from ultra-clean VAR/EBR Nitinol tubing used in the HighLife Mitral Valve Replacement system and tested under physiologically relevant conditions to 107, 108, and 4 × 108 cycles. Testing included multiple combinations of mean strains (1.5%-9%) and strain amplitudes (0.50%-2.50%) to simulate the multi-strain operating environments encountered in TMVR devices. VAR/EBR Nitinol demonstrated a conservative 130% improvement in 4 × 108-cycle Fatigue Strain Limit (FSL) compared to conventional VAR Nitinol at mean strains between 3% and 5%. The FSL behavior revealed two distinct strain regimes correlating with stress-induced martensitic transformation. Fractographic analysis confirmed elimination of inclusion-initiated fatigue failure, with crack initiation occurring independently of microstructural defects. The ultra-clean microstructure of VAR/EBR Nitinol enables a fundamental shift from flaw-dominated to stress-dominated fatigue behavior, providing unprecedented safety margins for complex transcatheter devices operating across diverse mechanical conditions. This material advancement has broad implications for Class III cardiovascular device design, enabling devices like the HighLife system and other TMVR platforms to meet stringent regulatory durability requirements while maintaining safety across complex multi-strain operating environments.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Design Example: Application of Archimedes' Principle
The volume of seawater displaced by the block is determined by first calculating the block's weight. This is done by multiplying the...
Fatigue
Anatomy of the Heart
Group Design
Muscle Recovery and Fatigue