Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Ultrasound Imaging of Cortical Bone: Cortex Geometry and Measurement of Porosity Based on Wave Speed for Bone
Amadou Sall Dia1, Guillaume Renaud2, Christine Chappard1
1Sorbonne Université, INSERM, CNRS, Laboratoire d'Imagerie Biomédicale, LIB, Paris, France.
Background:
It has been suggested that ultrasound (US) imaging can be used to assess cortical bone health, which is of particular interest owing to its major role in bone mechanical stability. Intra-cortical US imaging extends B-mode imaging into bone using a dedicated image reconstruction algorithm that corrects for refraction at the bone-soft tissue interfaces. It has shown promising results in a few healthy, predominantly young adults, providing anatomical images of the cortex (periosteal and endosteal surfaces) along with estimations of US wave speed. However, its reliability in older or osteoporotic bones remains uncertain.
Objective:
In this study, we critically assessed the performance of intra-cortical US imaging ex vivo in bones with various microstructural patterns, including bones exhibiting signs of unbalanced intra-cortical remodeling.
Methods:
We analyzed factors influencing US image quality, particularly endosteal surface reconstruction, as well as the accuracy of wave speed estimation and its relationship with porosity. We imaged 20 regions of interest from the femoral diaphysis of 5 elderly donors using a 2.5 MHz US transducer. The reconstructed US images were compared to site-matched high-resolution micro-computed tomography images.
Results:
In samples with moderate porosity, the endosteal surface was accurately identified, and thickness estimates from US and high-resolution micro-computed tomography differed by less than 10%. In highly remodeled bones with increased porosity, pore size and an heterogeneous distribution of pores, the reconstructed endosteal surface appeared less bright and was located above the trabecularized cortex region. We observed a decrease in US wave speed with increasing cortical porosity, aligning well with literature data, suggesting that, based on wave speed value the method could discriminate between bones with low porosity (<5%) and those with moderate to high porosity (>10%).
Conclusion:
This study paves the way for the application of US imaging in diagnosing cortical bone health, particularly for detecting increased cortical porosity and reduced cortical thickness.
Related Concept Videos
Bone Remodeling
Osteoclasts in Bone Remodeling
Speed of a Transverse Wave
One of the key properties of any wave is the wave speed. Light...
Propagation Speed of Electromagnetic Waves
Coordination Number and Geometry
Spongy Bone
Spongy bone is more porous, and less dense compared to compact bone. It is composed of concentric lamellae that are arranged irregularly to form the trabecular network. In some bones, the spaces between trabeculae contain red marrow, where...