Image-Guided Microwave Ablation of Liver Tumours Using a 150-W Generator: Ex Vivo and Multicenter in Vivo
Marco Calandri1,2, Susan Van Der Lei3,4, Valentina Cignini5
1Department of Surgical Sciences, University of Turin, Via Genova 3, 10126, Turin, Italy.
Purpose:
To assess the predictability of Emprint HP-Generator (150 W, 2.45 GHz) in terms of ablation volume and sphericity both in the ex vivo and in vivo setting.
Materials And Methods:
In an ex vivo setting, 48 ablations were performed on porcine livers at varying ablation times with a constant power of 150 W. MRI assessment of the ablation volumes was performed. In the in vivo setting consecutive procedures with 150W application with no rampage, single antenna positioning and availability of a contrast-enhanced computed tomography (CECT) immediately after the ablation were included. A semi-automated segmentation tool was used for volume estimation and spherical index (SI) was calculated.
Results:
In the ex vivo group, mean ablation time was 5.3 ± 2.5 min (range 1-10 min), mean ablation diameter was 37.7 ± 5.8 mm (range 21.3-49 mm) and mean ablation volume was 20.1 ± 10.8 cm3 (range 3-38.4 cm3) with a mean SI of 0.83 ± 0.08. In the in vivo group 78 percutaneous liver ablations (tumour mean diameter 23.2 ± 11 mm) in 65 patients (M46; F19; mean age 66.8 ± 11.3) were retrospectively enrolled from two institutions. The mean ablation time was 4,9 ± 2,5 min (range 1-10 min), mean ablation diameter was 41.3 ± 9.6 mm (range 16-73 mm) and mean ablation volume was 27.8 ± 15.6 cm3(range 6.4-105.6 cm3). Mean SI was 0.85 ± 0.11. Ablation volumes tend to be larger in case of hepatocellular carcinoma (HCC) compared to other histology (33.7 ± 21.2 cm3 vs 24.8 ± 10.5 cm3; p = 0.08). Both in ex vivo and in vivo settings, the SI remained stable regardless of ablation time and volume (Pearson's r 0.08, p 0.46), while ablation volumes increased linearly with time (Pearson's r 0.79, p < 0.001). Clavien-Dindo 1 complications were registered (4/65, 6.1%).
Conclusion:
The Emprint HP-generator enables the creation of large predictable volumes in both settings. SI remains consistent regardless of ablation time and volume.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
09:29Technique of Subnormothermic Ex Vivo Liver Perfusion for the Storage, Assessment, and Repair of Marginal Liver Grafts
Published on: August 13, 2014
Related Concept Videos
Predicting Molecular Geometry
Equivalence: In Vitro and In Vivo Bioequivalence
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test
Prediction Intervals
However, the point estimate is most likely not the exact value of the population parameter, but close to it. After calculating point estimates, we construct interval estimates, called confidence intervals or prediction intervals. This prediction interval comprises a range of values unlike the point estimate and is a better predictor of the observed sample value, y.
End Point Prediction: Gran Plot
For potentiometric titration, the Gran plot is created by plotting...
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...