Recanalization of Thrombosed Atrial Flow Regulator in Failing Fontan: A Bridge to Transplant
Raymond N Haddad1,2,3,4, Estibaliz Valdeolmillos1,2,3, Emre Belli1
1Filière des Cardiopathies Congénitales Enfant Adultes, Hôpital Marie Lannelongue, Centre Constitutif du Réseau Maladies Rares Malformations Cardiaques Congénitales Complexes - M3C, Hôpitaux Saint Joseph et Marie Lannelongue, Le Plessis-Robinson, France.
Abstract:
We report the case of a 10-year-old boy (41 kg, 133 cm) with hypoplastic left heart syndrome (mitral and aortic atresia), who underwent staged palliation: Norwood-Sano at 6 days, Glenn at 7 months, and Fontan completion at 3 years using a 16 mm non-fenestrated extracardiac Gore-Tex conduit. By age 8, he developed Fontan failure with elevated pressures and protein-losing enteropathy, refractory to medical management, supportive therapy, and hepatic lymphatic embolization. At age 9, the Fontan conduit was needle-perforated, dilated with a 7 mm balloon, and a 6 × 5 mm Occlutech atrial flow regulator (AFR) was implanted. Complete AFR thrombosis occurred 4 months later. Recanalization was achieved using an Astato® 30 0.018" wire under carotid protection with 6 mm SpiderFX filters, followed by high-pressure ballooning and a 9 × 18 mm BeSmooth stent implantation. Despite a patent AFR and optimized therapy, the patient showed no clinical improvement and ultimately underwent successful heart transplantation 1 month later under more favorable conditions.
More Related Videos
16:19Heterotopic Auxiliary Rat Liver Transplantation With Flow-regulated Portal Vein Arterialization in Acute Hepatic Failure
Published on: September 13, 2014
08:05Transplantation of Schwann Cells Inside PVDF-TrFE Conduits to Bridge Transected Rat Spinal Cord Stumps to Promote Axon Regeneration Across the Gap
Published on: November 3, 2017
Related Concept Videos
Design Example: Strain Gauge Bridge or Wheatstone Bridge
Bridge rectifier
Operationally, the bridge rectifier allows current flow through two of its diodes during each...
Wheatstone Bridge
Thus, for accurate resistance measurements, a...
Cross-bridge Cycle
Epigenetic Regulation
GTPases and their Regulation
Large G-proteins,...