Non-invasive analysis of pump parameter responses to orthostatic transitions in patients with fully magnetically
Lukas Ruoff1,2, Gregor Widhalm1, Michael Röhrich3
1Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Waehringer Guertel 18-20, Vienna 1090, Austria.
Aims:
Despite the excellent clinical outcomes of the HeartMate 3 (HM3) left ventricular assist device, the current pump monitoring limits in-depth pump data analysis. This study investigated HM3 pump parameters collected non-invasively with HM3 Snoopy during orthostatic transitions (OTs).
Methods And Results:
In this single-centre cohort study, a standardized OT protocol was developed, involving postural changes between supine, seated, and standing. Data were recorded using the HM3 Snoopy and a Holter electrocardiogram. Pump flows (QMIN, QMEAN, QMAX), pulsatility index (PI), pump speed, MagLev parameters, and heart rate were synchronized per second. The primary outcome was the identification of distinct orthostatic pump flow response phenotypes. Further, a binary classifier using MagLev parameters, to differentiate between supine and upright patient positions, was developed and assessed. In 25 HM3 patients (age: 61.2 ± 9.6 years, female: 12%, body mass index: 26.8 ± 4.7 kg/m2), greater flow alterations were observed during transitions from supine to standing vs. seated to standing, with most significant changes in QMIN [3 (-13; 10)%]. Phenotypes were identified across 75 OTs as no flow response (60%), undesired unloading with a loss in QMIN ≥ 50% (20%), and loss of pulsatility ≥ 50% (16%). The MagLev patient position classifier achieved a median sensitivity of 88% and specificity of 86% across the entire cohort.
Conclusion:
Three HM3 pump flow response phenotypes were identified in response to OTs, challenging the utilization of PI events to detect undesired unloading events. A MagLev-based position classifier revealed potential for differentiation of HM3 patient position.
Related Concept Videos
Properties of Transition Metals
Phase Transitions
Noncompartmental Analysis: Miscellaneous Pharmacokinetic Parameters
One key aspect of the noncompartmental approach is determining a drug's total clearance. This can be done by dividing the drug dose by the area under the concentration-time curve from zero to infinity. The area under the concentration-time curve represents the drug's...
Cooperative Allosteric Transitions
Noncompartmental Analysis: Mean Transit, Absorption and Dissolution Time
One of the key parameters is the mean transit time (MTT), which refers to the total duration required for drug molecules to transit through the body. MTT is determined by calculating the ratio of the area under the moment curve to the area...
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...