Performance of the Fifth Generation of a 16-Day Continuous Glucose Monitoring System in Adults with Diabetes
Shichen Chang1, Yibing Lu2, Hongwei Ling3
1Department of Endocrinology and Metabolism, Zhongshan Hospital, Fudan University, Shanghai, China.
Background:
This study evaluated the accuracy and safety of the Anytime 5Pro continuous glucose monitoring (CGM) system, a real-time, factory-calibrated device, over a 16-day period in adults with diabetes.
Methods:
Adult participants with type 1 or type 2 diabetes were recruited from three clinical sites in China. Each participant was equipped with four sensors (one on each upper arm and two on the abdomen) for a period of up to 16 days. To evaluate sensor performance, participants were randomly assigned to one of three 7-hour clinic sessions on days 1 or 2, days 7, 8 or 9, and day 16. During the sessions, the real-time glucose values measured by Anytime 5Pro CGM system were compared with venous blood glucose values measured by the EKF blood glucose detector. Primary endpoints for assessment were the mean absolute relative difference (MARD), the proportion of CGM values within ±20%/±20 mg/dL of reference values, and the percentage of paired points within Zones A and B of the consensus error grids.
Results:
In the cohort of 72 participants (287 sensors), the CGM system exhibited an overall MARD of 8.58%, with 96.59% of values within the ±20%/±20 mg/dL criteria. Comparative analysis revealed similar accuracy between arm (MARD: 8.58%; agreement: 97.03%) and abdomen (MARD: 8.58%; agreement: 96.16%) sensor placements. Throughout the 16-day wear period, the ±20%/±20 mg/dL agreement rates remained above 95%, with MARDs ranging from 8.30%-8.88%. Consensus error grid analysis showed 99.99% of points in Zones A and B. No serious adverse events were reported.
Conclusions:
The system demonstrated accurate glucose measurements across the 16-day wear period, irrespective of sensor placement site or glucose concentration.
More Related Videos
09:56Real-time Monitoring of Reactions Performed Using Continuous-flow Processing: The Preparation of 3-Acetylcoumarin as an Example
Published on: November 18, 2015
09:15Come to the Light Side: In Vivo Monitoring of Pseudomonas aeruginosa Biofilm Infections in Chronic Wounds in a Diabetic Hairless Murine Model
Published on: October 10, 2017
Related Concept Videos
Cattell's 16 Personality Factors
In contrast, source traits are the...
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Pathophysiology of Diabetes
Type 1 diabetes is characterized by autoimmune-mediated destruction of pancreatic β cells, with environmental factors potentially triggering this process in genetically susceptible individuals. Despite many not having a family history, certain genes increase susceptibility,...
Glucose Transporters
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes:
Adult Stem Cells
Continuing Care