Design of a rigid subacromial implant to treat massive irreparable rotator cuff tears
Cole T Fleet1, Daniel S McNeil2, J Andrew I Trenholm3
1Department of Mechanical and Materials Engineering, Western University, 1151 Richmond St., London, Canada; Roth | McFarlane Hand and Upper Limb Centre, St Joseph's Health Care, 268 Grosvenor St., London, Canada.
Background:
Massive rotator cuff tears often lead to superior migration of the humeral head and abnormal shoulder kinematics and function. In an attempt to restore normal joint kinematics, a modular subacromial implant was developed and evaluated in this biomechanical cadaveric study.
Methods:
Eight cadaveric specimens were assessed using a muscle-loading shoulder simulator that applied loads at 0, 30, and 60 degrees of glenohumeral abduction. Testing states included the intact rotator cuff (native), the simulation of a massive rotator cuff tear, and multiple implant states. Two implant thicknesses (5 and 8 mm) and two implant articular surface constraints (low & high) were tested. Humeral head translation in both anterior-posterior and superior-inferior directions was measured using an optical tracking system.
Findings:
The presence of a massive rotator cuff tear resulted in significant superior humeral head migration relative to the native state (P = 0.016). Both low and high constrained 5 mm thick implants restored superior-inferior humeral head position to within 1 ± 2 mm (P = 0.223) and 1 ± 2 mm (P = 0.928) respectively, of the intact position. Significant inferior translation relative to the native state was observed with both 8 mm implant designs (8 mm low constrained: -4 ± 2 mm, P = 0.060; 8 mm high constrained: -4 ± 2 mm, P = 0.060). No significant differences were observed for anterior-posterior translation.
Interpretation:
These results demonstrate that an implant affixed to the acromion helps to restore the native position of the humeral head in the presence of a massive rotator cuff tear with superior migration. The 5 mm high constrained implant design was most effective at restoring native humeral head position.
Level Of Evidence:
Basic Science; Biomechanics.
Related Concept Videos
Rigid Body Equilibrium Problems - I
Rigid Body Equilibrium Problems - II
Consider two children sitting on a seesaw, which has negligible mass. The first child has a mass (m1) of 26 kg and sits at point A, which is 1.6 meters (r1) from the pivot point B; the second child has a mass (m2) of 32 kg and sits at point C. How far from the pivot point B should the second child sit (r2) to balance the seesaw?
Kinetic Energy for a Rigid Body
Equation of Motion for a Rigid Body
The combined moments generated about the center of mass of the object are equal to the rate of change of the angular momentum of the body. An external force, when applied at a different...
Virtual Work for a System of Connected Rigid Bodies
Next,...
Angular Momentum: Rigid Body
This calculation can get complicated when tiny particles within the rigid body are not rotating in the same plane but have...