Measuring fluorotelomer alcohols by thermal desorption-gas chromatography-tandem mass spectrometry: Interlaboratory
Clara M A Eichler1, Hannah Calder2, Bharat Chandramouli3
1Oak Ridge Institute for Science and Education Postdoctoral Fellow at the U.S. Environmental Protection Agency, Office of Research and Development, Center for Environmental Measurement and Modeling, Research Triangle Park, NC 27711, USA.
Abstract:
Fluorotelomer alcohols (FTOHs) are a group of volatile and semi-volatile per- and polyfluoroalkyl substances (PFAS) commonly found in indoor air and contribute to PFAS inhalation exposure. Improving our understanding of the contribution of FTOHs to human exposure to PFAS indoors is of high interest. Consensus standard test methods play an essential role in environmental risk assessment and management. We present the outcome of an international interlaboratory study (ILS) conducted with nine laboratories to evaluate the precision of ASTM International Standard Test Method D8591. The ILS was organized by the United States Environmental Protection Agency. The test method specifies the analysis of four FTOHs (4:2, 6:2, 8:2 and 10:2 FTOH) collected on PFAS-specific thermal desorption tubes by gas chromatography coupled with tandem mass spectrometry. During the ILS, participating laboratories were instructed to use the test method to analyze three samples (A, B, and C), each three times. Each sample contained the target FTOHs at defined concentrations unknown to the laboratories. The results from seven laboratories show that the relative reproducibility standard deviation (RSDR) of the method ranges from 14% to 26% and the relative repeatability standard deviation (RSDr) ranges from 4.6% to 11%, with RSDs decreasing with decreasing volatility of the FTOHs. Bias ranged from -13% to 6.0% and was generally larger and negative for less volatile FTOHs. The test method in conjunction with the precision statistics from this ILS will provide a reliable, defendable method that can be used in the context of studying PFAS sources, transport, and human exposure.
More Related Videos
08:56Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
08:14Extraction of Non-Protein Amino Acids from Cyanobacteria for Liquid Chromatography-Tandem Mass Spectrometry Analysis
Published on: December 9, 2022
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Tandem Mass Spectrometry
Mass Spectrometry: Alcohol Fragmentation
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...