Initial experience with the extravascular implantable cardioverter-defibrillator: Workflow, feasibility and safety
Ana Lousinha1, Hagen Kahlbau2, Sara Coelho3
1Serviço de Cardiologia, Unidade de Arritmologia, Pacing e Eletrofisiologia, Hospital de Santa Marta, Centro Hospitalar Universitário de Lisboa Central, EPE, Lisboa, Portugal; Hospital CUF Descobertas, Lisboa, Portugal; NOVA Medical School, Faculdade de Ciências Médicas, NMS, FCM, Universidade NOVA de Lisboa, Lisboa, Portugal.
Introduction And Objectives:
Implantable cardioverter-defibrillators (ICDs) are among the most effective interventions for the prevention of sudden cardiac death. However, traditional transvenous systems are associated with venous access and intracardiac lead complications. Subcutaneous ICDs (S-ICD), which were developed to overcome these complications, are unable to provide anti-tachycardia pacing (ATP) or pause prevention pacing. The extravascular ICD (EV-ICD) is a recent technological innovation designed to mitigate these limitations by enabling cardioversion, defibrillation and pacing without entering the central venous system. Our aim was to assess the initial experience with the new EV-ICD system, as a valid alternative to transvenous ICD and S-ICDs, including patient selection, implantation technique, procedure feasibility and safety, functional parameters, and immediate clinical outcomes.
Methods:
We conducted an observational study of all patients who underwent EV-ICD implantation between November 2024 and June 2025 at two centers. Demographic, clinical, imaging, pharmacologic and procedural data were analyzed.
Results:
A total of 11 patients were included, with a mean age of 36.3 years (range 19-59); 27% were female. Main diagnoses included hypertrophic cardiomyopathy (n=3), left ventricular non-dilated cardiomyopathy (n=3), dilated cardiomyopathy (n=2), Brugada syndrome (n=1), polymorphic ventricular tachycardia (n=1) and arrhythmogenic right ventricular cardiomyopathy (n=1). There were no procedural or peri-procedural complications. The mean procedure duration ("skin-to-skin") was 68.2 minutes (range 60-78). Median fluoroscopy time was 3.9 minutes (range 2.5-6.2). Defibrillation threshold (DFT) testing was successful in all cases. During the follow-up, two patients received an inappropriate shock, one due to sinus tachycardia and the other due to oversensing of myopotentials.
Conclusions:
Our initial experience with the EV-ICD confirms the feasibility and safety of the procedure. This novel system may offer an effective alternative for selected populations, especially young patients with extended life expectancy and high cumulative device exposure. The rate of inappropriate shocks remains a concern and improvements are required to reduce these events.
More Related Videos
13:54A Workflow for Lipid Nanoparticle LNP Formulation Optimization using Designed Mixture-Process Experiments and Self-Validated Ensemble Models SVEM
Published on: August 18, 2023
12:47Workflow Based on the Combination of Isotopic Tracer Experiments to Investigate Microbial Metabolism of Multiple Nutrient Sources
Published on: January 22, 2018
Related Concept Videos
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Two-Compartment Open Model: Extravascular Administration
The absorption exponent (ka) indicates the speed at which the drug...
Survey Safety
One-Compartment Open Model for Extravascular Administration: First-Order Absorption Model
Non-Oral Extravascular Drug Absorption Routes
Lipophilic drugs that are stable at salivary pH (6) and exhibit minimal binding to the oral mucosa are absorbed more...