Related Experiment Video
Updated: Jan 27, 2026

Lower Limb Biomechanical Analysis of Healthy Participants
Published on: April 15, 2020
Structural design and biomechanical analysis of anti- subsidence gradient porous artificial vertebrae combined with a
Peng Shang1,2, Xinyu Wu3, Rongqian Li3
1Hebei University of Technology, Tianjin, 300400, China. shangpeng@hebut.edu.cn.
Background:
In total disc replacement, mechanical failure and implant subsidence are significant complications after vertebral body replacement. Therefore, the design of a new artificial vertebral implant(AVI)that can ensure safety and improve prosthesis's subsidence has become a key challenge for spinal reconstruction.
Methods:
This study involved designing a new AVI. Its exterior is a topological entity framework, and its interior is a gradient triply periodic minimal surface (TPMS) gyroid structure, which simulates cortical bone and cancellous bone, respectively. To enhance the safety and stability of the prosthesis. The biomechanical properties of AVIs under different physiological postures were analysed using the finite element method(FEM). The samples were prepared using selective laser melting (SLM), and the quasi-static compression and subsidence experiments were carried out for evaluation.
Results:
The peak stress of the topological gradient gyroid AVI was significantly lower than that of the traditional titanium mesh cage (TMC) and the truss + fluorite AVI, with a maximum reduction of 69.7%. Under all working conditions, the peak stress at the bone-implant interface of the vertebrae was 5.9% -14.1% lower than that of the truss + fluorite AVI and much lower than that of the TMC, with a decrease of 44.3%. Quasi-static compression experiments show that the maximum compression load of the topological gradient gyroid AVI is approximately 1.12 times greater than that of the truss + fluorite AVI and significantly greater than that of the TMC. This indicates that topological gradient gyroid AVI meets safety requirements. Additionally, its anti-subsidence ability is 20.4% higher than that of traditional porous AVI.
Conclusions:
Compared with traditional AVIs, the combination of topology optimization framework and gradient gyroid structure significantly improves the mechanical strength and subsidence resistance of the AVI, and can effectively reduce the stress concentration, which is expected to become the preferred scheme for spinal reconstruction.
Related Concept Videos
General Structure of a Vertebra
Structure-Activity Relationships and Drug Design
SAR studies the intricate relationship between a drug's chemical structure and biological activity. It focuses on understanding how modifications to a drug's structure can influence...
What is an Electrochemical Gradient?
The chemical gradient relies on differences in the abundance of a substance on the outside versus the inside of a cell and flows from areas of high to low ion concentration. In contrast, the electrical gradient revolves around an...
Group Design
Factorial Design
Design of Transmission Shafts - Stress Analysis