Optimized Management of Endovascular Treatment for Acute Ischemic Stroke
Published on: January 18, 2018
Value of Thrombus Imaging Characteristics as a Guide for First-Line Endovascular Thrombectomy Device in Patients With
Nikki Boodt1,2,3, Agnetha A E Bruggeman4, Manon Kappelhof4,5
1Department of Radiology and Nuclear Medicine Erasmus MC, University Medical Center Rotterdam Rotterdam The Netherlands.
Background:
It has been suggested that selection of a first-line endovascular thrombectomy device, that is, contact aspiration (CA) or stent retriever (SR) thrombectomy, could be based on thrombus type. Thrombus composition and mechanical behavior can partially be predicted with thrombus computed tomography (CT) characteristics. We aimed to assess the influence of thrombus CT characteristics on the association between first-line device and outcomes of endovascular thrombectomy.
Methods:
For patients enrolled in the MR CLEAN (Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in The Netherlands) Registry between March 2014 and November 2017, we assessed thrombus density, thrombus length, and presence of hyperdense artery sign on thin-slice (≤2.5 mm) admission CT. We used regression models to estimate the relationship between first-line endovascular thrombectomy device (CA versus stent retriever) and first-pass reperfusion (FPR, expanded Thrombolysis in Cerebral Infarction score 2C-3 after first attempt), final reperfusion, procedure duration, 24-hour National Institutes of Health Stroke Scale, and 90-day modified Rankin scale score and tested for interaction of thrombus characteristics with first-line device by adding interaction terms.
Results:
Of 703 included patients, 520 (74%) received first-line stent retriever and 183 (26%) first-line CA. Overall, the first-line device was not associated with FPR (adjusted odds ratio [aOR], 1.32 [95% CI, 0.88-1.98]). In patients with thrombus density below the median (<50 Hounsfield units), FPR was more often achieved with CA than with a stent retriever (34% versus 24%, aOR, 1.95 [95% CI, 1.09-3.50]), whereas in patients with thrombus density above the median (≥50 Hounsfield units), first-line device was not associated with FPR (aOR, 0.90 [95% CI, 0.50-1.62]). The interaction between thrombus density as a continuous variable and first-line device on outcome was not significant (P=0.38). There was also no interaction between first-line device and the other thrombus characteristics for FPR or the other outcomes.
Conclusion:
Our study does not provide evidence that the association between first-line thrombectomy device and endovascular thrombectomy outcomes depends on thrombus CT characteristics. Based on our results, there are no arguments for thrombectomy device selection based on thrombus CT characteristics. A possible better performance of CA in low-density, fibrin-rich clots needs further study.
Related Concept Videos
Characteristics of Life
Characteristics of Fluids
Characteristics of Fluids
Fluids, which include both liquids and gases, are substances that deform continuously under shearing stress. For example, water and oil are liquids with molecules that can...
Characteristics of BJT
For input characteristics, the base-emitter voltage is varied, maintaining a constant collector-emitter voltage. This setup reveals a Shockley-type dependence of the collector current on...
Characteristics of JFET
The core of a JFET's operation is controlling drain current by modulating the gate-source voltage. When the drain and gate voltage are set to zero, the JFET exhibits no net current flow, representing a state of equilibrium. The drain current increases linearly as the...
Characteristics of OpAmp