Improvement in positional accuracy with integrated surface- and X-ray imaging for intracranial stereotactic
Caisa Kjellström1,2, Tobias Pommer1,2, Peter Siesjö3
1Medical Radiation Physics, Lund University, Lund, Sweden.
Background And Purpose:
Stereotactic radiosurgery (SRS) requires high positional accuracy to safely deliver large doses. This study evaluated an integrated surface- and image-guided radiotherapy (SGRT-IGRT) system by analysing (1) the agreement between thermo-optical and stereoscopic X-ray positioning, and (2) the impact of intra-fractional workflows on treatment accuracy and time.
Materials And Methods:
Data from 126 SRS patients treated with 30 Gy/3 fractions (n = 116) or 12 Gy/1 fraction (n = 10) on a Varian Truebeam STx were retrospectively analysed. Patients were positioned and monitored with Brainlab ExacTrac Dynamic, with 0.5 mm/0.5° tolerances for IGRT and 1 mm/1° for SGRT. Three workflows were investigated: (A) SGRT + IntraArc IGRT (imaging every 90° during treatment and between couch rotations); (B) SGRT + InterArc IGRT (imaging between couch rotations only); and (C) SGRT (no additional imaging after initial coplanar setup). Workflows (B) and (C) were simulated by omitting applied couch corrections.
Results:
Median beam-on times were 5.5 min for workflow A, 5.0 min for workflow B, and 3.2 min for workflow C. The median differences between thermo-optical and stereoscopic X-ray patient positioning were ≤0.1 mm. The 3D positioning uncertainty remained within 0.5 mm (2.5th-97.5th percentile) using SGRT-IGRT. Omitting inter-arc imaging increased positional deviation ranges from 0.1-0.5 mm to 0.1-0.7 mm.
Conclusion:
Thermo-optical and stereoscopic X-ray imaging showed good agreement within the set institutional tolerances. Inter-arc imaging increased treatment time by 2 min compared with SGRT alone but improved positioning accuracy. Intra-arc imaging added an additional small accuracy benefit at minor time cost.
More Related Videos
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Velocity and Position by Integral Method
Consider an example to calculate the velocity and position from the acceleration function. A motorboat is traveling at a constant velocity of 5.0 m/s when it starts to decelerate to arrive at the dock. Its acceleration is...
Line, Surface, and Volume Integrals
X-ray Imaging
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...