Comparative Dosimetry of Handheld Versus Wall-Mounted Intraoral X-ray Systems: A Phantom-Based Study
Vignesh Guptha Raju1, Rajsandeep Singh2, Ameer Akhil Ahmed Shaik3
1Department of Pedodontics and Preventive Dentistry, Karpaga Vinayaga Institute of Dental Sciences, Chennai, IND.
Introduction:
The rapid adoption of handheld dental X-ray units has revolutionized imaging in mobility-restricted patients; however, concerns may still persist regarding their radiation safety profile compared with traditional wall-mounted systems. The aim of the present study was to quantitatively compare operator- and patient-equivalent radiation doses, ambient scatter, and diagnostic image quality between a handheld (MaxRay Cocoon, Dexcowin Co. Ltd., Korea) and a wall-mounted (Optima DC, Alerio Technologies Pvt. Ltd., India) intraoral X-ray unit during standardized mandibular molar radiography.
Materials And Methods:
This prospective experimental study was conducted using a human cadaver mandible as a standardized phantom. A total of 384 exposures (198 per device) were performed with identical exposure time (0.18 seconds), rectangular collimation, and a size-2 complementary metal-oxide-semiconductor (CMOS) digital sensor using the paralleling technique. Operator doses were recorded using real-time electronic personal dosimeters (RaySafe i3, RaySafe, Sweden) at the chest and thyroid levels and thermoluminescent dosimeters (TLD-100) placed under the lead apron at the chest, forehead, and fingers of the dominant hand. The patient-equivalent effective dose was calculated from dose-area product (DAP) measurements using conversion coefficients recommended by the International Commission on Radiological Protection. Image quality was evaluated objectively through signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) measurements and subjectively by three blinded experienced oral radiologists using a five-point Likert scale. Data were analyzed using independent sample t-tests.
Results:
The handheld unit produced significantly higher operator doses: finger dose 0.45 ± 0.12 µSv vs. 0.08 ± 0.03 µSv (p < 0.001), chest (outside apron) 0.18 ± 0.05 µSv vs. 0.05 ± 0.02 µSv (p = 0.001), and thyroid 0.11 ± 0.03 µSv vs. 0.04 ± 0.01 µSv (p = 0.003). The effective dose to the mandible phantom was 2.65 ± 0.25 µSv (handheld) vs. 1.82 ± 0.18 µSv (wall-mounted) (p < 0.001). The ambient scatter at 1 m was three times higher for the handheld device (p < 0.001). No significant differences were found in SNR (p = 0.125), CNR (p = 0.08), or subjective diagnostic quality scores (p = 0.65).
Conclusions:
Despite comparable diagnostic image quality, the handheld dental X-ray unit delivered substantially higher radiation doses to both the operator and the patient-equivalent phantom than the wall-mounted system. Wall-mounted units should remain the standard where patient mobility permits, reserving handheld systems for clinically justified scenarios.
Related Concept Videos
Comparative Excretory Systems
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Plant Cell Wall
X-ray Imaging
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...