Low-Dose CT Quality Assurance at Scale: Automated Detection of Overscanning, Underscanning, and Image Noise
Patrick Wienholt1,2, Alexander Hermans1,2,3, Robert Siepmann1,2
1Lab for Artificial Intelligence in Medicine, Department of Diagnostic and Interventional Radiology, University Hospital RWTH Aachen, 52074 Aachen, Germany.
Abstract:
Automated quality assurance is essential for low-dose computed tomography (LDCT) lung screening, yet manual checks strain clinical workflows. We present a fully automated artificial intelligence tool that quantifies scan coverage and image noise in LDCT without user input. Lungs and the aorta are segmented to measure cranial/caudal over- and underscanning, and noise is computed as the standard deviation of Hounsfield units (HUs) within descending aortic blood, normalized to a 1 mm3 voxel. Performance was verified in a reader study of 98 LDCT scans from the National Lung Screening Trial (NLST), and then applied to 38,834 NLST scans reconstructed with a standard kernel. In the reader study, lung masks were rated ≥"Nearly Perfect" in 90.8% and aorta-blood masks in 96.9% of cases. Across 38,834 scans, mean overscanning distances were 31.21 mm caudally and 14.54 mm cranially; underscanning occurred in 4.36% (caudal) and 0.89% (cranial). The tool enables objective, large-scale monitoring of LDCT quality-reducing routine manual workload through exception-based human oversight, flagging protocol deviations, and supporting cross-center benchmarking-and may facilitate dose optimization by reducing systematic over- and underscanning.
More Related Videos
Related Concept Videos
Quality Assurance
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
pH Scale
Dose Size and Dosing Frequency: Determination Methods