Background:
International robotic-assisted PCI platforms such as CorPath 200 and CorPath GRX have demonstrated procedural safety and efficacy but are limited by the absence of real-time tactile feedback and cumbersome device exchange. The ETcath200, the first domestically developed robotic PCI system in China, introduces guidewire force-sensing to provide real-time visual resistance feedback that partially compensates for the lack of direct tactile sensation, along with streamlined device exchange and an intuitive joystick-roller interface.
Aims:
This study aimed to evaluate the feasibility, safety, and procedural performance of the ETcath200 system in patients with complex coronary lesions.
Methods:
In this prospective, single-center, single-arm trial, 20 patients with AHA/ACC type B2 or C complex coronary lesions underwent PCI using the ETcath200 system. Technical success was defined as full robotic completion without any unplanned manual input (including partial manual assistance or manual conversion); clinical success as residual stenosis < 30% with TIMI 3 flow and no in-hospital major adverse cardiac events (MACE). Patients were followed for 30 days.
Results:
All 20 patients (mean age 63.2 ± 12.0 years; 80% male) had unstable angina and complex coronary anatomy (79.2% type C lesions; 50% bifurcation, 65% diffuse). Most procedures were performed via radial access (80%), and the majority utilized drug-eluting stents (65%). Technical success was achieved in 16 cases (80%), while four cases (20%) required manual conversion due to severe tortuosity or calcification (n = 3) or guidewire-induced dissection (n = 1; managed successfully). All patients achieved clinical success; no device failures, MACE, or serious adverse events occurred at 30 days. Mean procedure duration was 61.0 ± 23.3 min; radiation dose was 0.07 ± 0.04 mSv.
Conclusions:
The ETcath200 robotic PCI system demonstrated feasibility, safety, and reliable performance in complex coronary lesions, achieving high clinical success and excellent short-term outcomes. Manual conversion was required only in anatomically challenging cases. These data support further multicenter evaluation of ETcath200 in complex coronary interventions.
More Related Videos
11:03Robotic Left Hepatectomy using Indocyanine Green Fluorescence Imaging for an Intrahepatic Complex Biliary Cyst
Published on: June 24, 2022
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
Related Concept Videos
Survey Safety
Acute Coronary Syndrome III: Diagnostic Studies
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Center of Gravity
Center of Gravity
To determine its location, the principle of moments can be utilized by dividing the object into...
Center of Mass
The knowledge of the center of mass can also help us to describe and predict the motion of objects. For example, when a ball is thrown...