Comparison of two swept-source optical biometers in eyes with long axial length
Pedro Tañá-Rivero1, Paz Orts-Vila1, Pedro Tañá-Sanz1
1Oftalvist, Alicante, Spain.
Purpose:
To assess the repeatability and agreement of two swept-source optical coherence tomography (SS-OCT)-based biometers when measuring biometric parameters in cataract eyes with a long axial length.
Design:
Cross-sectional prospective, comparative study.
Methods:
One-hundred and twenty-six eyes of 126 patients with cataracts (LOCS III grade equal or less than 3) and axial lengths >25 mm were measured 3 consecutive times using the Argos and IOLMaster 700 SS-OCT optical biometers. Keratometry [K1 (flat) and K2 (steep)], central corneal thickness (CCT), white-to-white (WTW), anterior-chamber-depth (ACD), lens-thickness (LT), and axial length were measured using both instruments. The repeatability for each device was analysed with the within-subject standard deviation (Sw), coefficient of variability, and coefficient of repeatability, and the agreement between the devices was analysed using a Bland-Altman plot.
Results:
The acquisition success rate for both biometers was 100%. For K1 and K2, the Sw values were <0.15D and <0.2D, respectively using the two instruments. For CCT, WTW, ACD and LT, the values were 6.56 μm, 0.05 mm, 0.01 mm and 0.01 mm with the Argos, and 3.54 μm, 0.07 mm, 0.01 mm and 0.01 mm with the IOLMaster 700. The Sw values for axial length were 0.01 and 0.03 mm for the Argos and IOLMaster 700, respectively. There were statistically significant differences between the two SS-OCT biometers for all the parameters evaluated (p < 0.001) except WTW (p = 0.088). The mean differences for K1, K2, CCT, WTW, ACD, LT and axial length were 0.102D, 0.133D, -14.224 μm, -0.018 mm, 0.108 mm, 0.029 mm, and -0.102 mm, respectively.
Conclusion:
Both SS-OCT biometers provide repeatable measurements for the different parameters analysed, however, when compared, it is clear that some of them must be assessed carefully in order to be considered interchangeable.
More Related Videos
03:47Retinal and Choroidal Thickness Changes in Populations with Helicobacter pylori Infection by Swept-Source Optical Coherence Tomography
Published on: November 1, 2024
03:35Determining Gender-Based Differences in Retinal and Choroidal Thickness in Underweight Individuals via Swept-Source Optical Coherence Tomography
Published on: December 1, 2023
Related Concept Videos
Sources of Self-Esteem III: Social Comparison
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Axial and Appendicular Muscles
Axial Muscles
Axial muscles, situated along the body's midline, are intricately connected to the axial skeleton, which includes the skull, spine, ribs, and sternum. These muscles facilitate facial expressions and...
Overview of the Axial Skeleton
The axial skeleton of the...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Normal Strain under Axial Loading