Assessment of bioelectrical impedance analysis devices for data reliability of body impedance measurements
Dan Bee Kim1, Seong Su Shin1, Wan-Seop Kim1
1Korea Research Institute of Standards and Science, Daejeon, Republic of Korea.
Background:
Bioelectrical impedance analysis (BIA) devices are widely used for body composition analyses, but their accuracy for body impedance measurements can vary significantly between BIA device models. Calibration of BIA device is essential to ensure the reliability of measurement data, particularly for clinical and research purposes. This study aims to evaluate the electric impedance measurement abilities of multi-frequency BIA devices using a home-built body impedance imitator (BII), designed to simulate human body impedance, and to assess the measurement uncertainty based on calibration with internationally accredited reference standards.
Methods:
Two different models of multi-frequency BIA device with similar specifications and measurement schemes were evaluated using the BII as a reference standard. First, the impedance values of the BII were calibrated using an LCR meter, traceable to national impedance standards. Next, the impedance measurements were performed using both models based on the standard BII over a wide frequency range (1 kHz to 1 MHz) to analyze the accuracy, linearity, contact resistance effects, etc. of the BIA devices.
Results:
Despite having similar specifications, two models of BIA device exhibited notable differences in their measurement performance. Model A showed better accuracy and consistency, with measurement uncertainty of approximately 1.4% at a 95% confidence level (k = 2). In contrast, Model B showed higher deviations, particularly in trunk impedance measurements. The study found that the BIA device calibration procedure could effectively quantify the measurement uncertainty and ensure the reliability of measurement data provided by the BIA device.
Conclusion:
This study demonstrates that BIA devices can be calibrated using a standard procedure with a stable reference such as the BII, which traces back to internationally accredited impedance standards. The results show that the calibration significantly improves the reliability of BIA device measurements, providing a level of confidence for the measurement data.
More Related Videos
07:44Author Spotlight: Implementation of BIVA for Analyzing Disease Risk Factors in Patients with Low Body Cell Mass
Published on: July 14, 2023
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
Related Concept Videos
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Series Impedances: Three-Phase Line
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...
Bus Impedance Matrix
In the first circuit, all machine voltage sources are short-circuited, leaving only the prefault voltage source at the fault location. The positive-sequence bus impedance matrix can be determined by solving the nodal equations,...
Line Protection with Impedance Relays
Under normal conditions, low load currents keep the measured...
RLC Series Circuits: Impedance
Thus, the magnitude of the impedance is given by the following equation,