Diagnostic accuracy of the HUAWEI WATCH D for measuring spinal anesthesia-induced hypotension during elective cesarean section: a diagnostic test
Related Experiment Video
Updated: Jan 31, 2026

Diagnostic Ultrasound Imaging of Mouse Diaphragm Function
Published on: April 21, 2014
Wei Li1, Yan Huan1, Xianjin Zhou1
1Shanghai Key Laboratory of Maternal-Fetal Medicine, Department of Anesthesiology, Shanghai Institute of Maternal-Fetal Medicine and Gynecologic Oncology, Shanghai First Maternity and Infant Hospital, School of Medicine, Tongji University, Shanghai, China.
Background:
The newly developed HUAWEI WATCH D (HW) is a wearable smartwatch, primarily designed to measure non-invasive blood pressure (NIBP) at the wrist. Two recent studies have demonstrated that the accuracy of HW is the same as that of mercury sphygmomanometer for measuring BP in adults. This study aimed to validate the accuracy of the HW in hypotension monitoring.
Methods:
A total of 118 healthy parturients with singleton pregnancies undergoing elective cesarean section were included. BP was measured three times preoperatively on the left wrist using HW and on the right upper limb using an Omron BP monitor, with the respective mean values recorded as the baseline values in the supine position. After spinal anesthesia, bilateral BP was measured simultaneously by HW and a NIBP monitor at 1-min intervals, 10 times. The primary outcome was the Chi-squared test results comparing hypotension detected between HW and NIBP, as well as the area under the curve (AUC) for receiver operating characteristic (ROC), the Youden index, and the cut-off point of HW for measuring hypotension.
Results:
Overall, 1180 pairs of systolic blood pressure (SBP) were measured. Using an SBP of <80% baseline values of NIBP as the gold standard for hypotension, Youden Index = 0.62, the hypotension threshold for HW is set to 104 mmHg, with sensitivity of 79.2%, specificity of 82.0%, false positive rate of 18.0%, and false negative rate of 20.8%, and accuracy of 81.3%. The Chi-squared test yielded a Pearson γ2 = 350.4, Kappa = 0.53, P < 0.001 (two-tailed). The AUC of ROC was 0.87 (95% confidence interval: 0.85-0.90), P < 0.001 (two-tailed).
Conclusion:
The HUAWEI WATCH D demonstrates moderate diagnostic accuracy for detecting severe hypotension (SBP < 80% baseline) during caesarian section under spinal anesthesia, with a sensitivity of 79.2%, a specificity of 82.0%, and an AUC of 0.87.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Appendicitis-II: Diagnostic Studies and Management
Diagnosing Appendicitis
It requires a multifaceted approach, starting with a detailed physical examination to pinpoint the location and nature of the pain and identify any associated symptoms. Laboratory tests play a crucial role. A complete Blood Count (CBC) typically reveals leukocytosis (an increased number of...
Myasthenia Gravis: Diagnostic Tests
The edrophonium test is a diagnostic tool for myasthenia gravis. It involves...
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
Pericarditis II: Clinical Features and Diagnostic Tests
Atherosclerosis II: Clinical Manifestations and Diagnostic Tests